Chapter 2, Machine Part Replacement Reference
2-19
Remove the protective paper from the other piece of tape and attach it
in the same manner inline with the other hinge nut plate.
6. Reinstall the membrane panel onto the top cover.
7. Reinstall the decorative panel onto the top cover.
8. Cut double-sided tape (P/N 18142-00) to a length of 1½" ± ¼".
Adhere the tape to the underside of one of the EMI ferrite attenuators
(P/N 39475-00).
9. Sandwich the ribbon cable from the top cover assembly between the
attenuator with the tape and the attenuator (P/N 39475-00) without the
tape at approximately 4" ± ¼" from the end of the cable jacket.
Note:
The attenuator with the tape must be toward the card cage.
10. Install the two spring clips (P/N 47241-00) over each end as shown,
reference balloon #50.
Note:
The flat end of the spring clip must be toward the card cage.
11. Remove the protective paper from the tape on the attenuator and
adhere the attenuator assembly with the ribbon cable to the top of the
card cage.
Note:
Do not cover slots in top of card cage.
12. Apply purple Loctite to the threads of the two #6-32 x 3/8" Phillips
flat head screws (removed previously). Make sure the ribbon cable
from the cover assembly is centered in the deck body clamp and
secure the copper tape of the cable assembly under the strap clamp
(P/N 47133-02) and secure in place with the Loctited screws.
Note IMPORTANT:
The copper tape must be centered in the strap
clamp.
13. Reinstall cable connector into the backplane card.
14. Complete a Diagnostic and Functional Test according to the
procedure outlined in Chapter 3.