6-22
PCS2 Service Manual
CODE
TEXT MESSAGE
DESCRIPTION
SUGGESTED RESOLUTION
AIR IN DONOR
LINE 1
AIR IN DONOR
LINE 2
ARRAY BOUNDS
ERROR
DLAD 1 air signal
indicates air when
fluid is expected.
DLAD 2 air signal
indicates air when
fluid is expected.
An internal software
error has occurred,
where an item was
accessed beyond a
defined array.
1. Use air detector test in Diagnostics to confirm problem.
2. Check, repair, or replace all electrical connections
associated with the DLAD1 air detector.
3. If fluid is present in tubing and tubing is installed in the
air detector, measure the (+5 VDC fluid, 0 VDC air)
DLAD1 signal at P503 pin 2 on the backplane card. If no
signal, replace DLAD1 air detector.
4. Replace processor card. Using Utilities, calibrate all the
components under A/D, centrifuge, safety system, and
configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.
1. Use air detector test in Diagnostics to confirm the
problem.
2. Check, repair, or replace all electrical connections
associated with the DLAD2 air detector.
3. If fluid is present in tubing and tubing is installed in the
air detector, measure the (+5 VDC fluid, 0 VDC air)
DLAD2 signal at P503 pin 4 on the backplane card. If no
signal, replace DLAD2 air detector.
4. Replace processor card. Using Utilities, calibrate all the
components under A/D, centrifuge, safety system, and
configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.
1. Check +5 VDC power supply.
2. Replace processor card. Using Utilities, calibrate all the
components under A/D, centrifuge, safety system, and
configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.