Chapter 5, Installation and Configuration
5-5
3. Press the MODIFY PROGRAM key until the arrow points to Draw
Cycle.
4. Select Draw Cycle by pressing the DRAW key.
5. Set the LOW weight by using the YES key to increase and the NO
key to decrease the weight. Press the MODIFY PROGRAM key when
the desired LOW weight has been reached.
6. Set the MID weight by using the YES key and NO key. Press the
MODIFY PROGRAM key when the desired MID weight has been
reached.
7. Set the HIGH weight by using the YES key and NO key. Press the
MODIFY PROGRAM key when the desired HIGH weight has been
reached. The screen will then display the “Suction Compens. wt.”
setting; leave the setting at 1.
8. Press the MODIFY PROGRAM key to set the Plasma Unit of
Measurement to the desired setting. Press the YES key to select
Volume and the NO key to select “Weight” (The default value is
Weight). Proceed to Step 10 if “Weight” is selected.
9. Press the MODIFY PROGRAM key after selecting “Volume” to set
the Plasma Density [%]. Use the YES and NO key to set the Plasma
Density [%] in accordance to the center SOP (The default value is
1026).
10.Press the MODIFY PROGRAM key until Duration in Statistics is
displayed. Use the YES key to ENABLE and the NO key to DISABLE
(The default value is Enable). Selecting ENABLE will display the
procedure time in the Procedure Statistics Screen.
11.Press the STOP key to exit Draw Cycle Configuration.
12.Press the STOP key again to exit Configuration. Toggle the write
protect switch on the Processor Card when instructed, and then replace
the rear panel.
13. Exit Utilities by pressing the STOP key.
Bowl Configuration
1. Power on the PCS2 in the Utilities mode of operation by pressing and
holding the PLASMA key while switching the power on. Hold the
PLASMA key until the Utilities menu appears on the display.
2. Select Configuration on the Utilities menu by pressing the MODIFY
PROGRAM key until the arrow points to Configuration, then press
the DRAW key. Toggle the write protect switch on the Processor Card
when instructed (remove rear access panel).
3. Press the MODIFY PROGRAM key until the arrow points to HS
Bowl, then press the DRAW key
4. Press the MODIFY PROGRAM key until HS Bowl is displayed. Use
the Yes key to ENABLE and the NO key to DISABLE (The default is
Disable).