Chapter 4, Calibration
4-3
Condor Power Supply
Note:
Use the following calibration procedure if the unit has a Condor
power supply assembly installed (P/N 48895-00).
1. Measure the power supply voltages at P507 on the
b
ackplane Card
(see Figure 4-1). Use pin seven for DC common/ground.
Note: Pins 1 thru 7 are all common/ground.
2. To calibrate the power supply voltages, adjust the +5V trim pot on the
front panel of the power supply. All other voltages will track the 5 volt
power supply. Although no adjustments can be made to the other
voltages, please check and verify that they are within their respective
specifications as described in Figure 4-1.
1. Power on the PCS2 in the Diagnostic mode of operation. The PCS2
must be powered on for a minimum of 15 consecutive minutes to
stabilize the optic circuit before proceeding with calibration. When 15
minutes have elapsed, remove the rear panel.
2. Inspect the optic window for contamination and clean it before
measuring the optics signal.
3. Remove the gray card from the envelope in the Gray Card
Replacement Kit.
4. Make a note of the Correction Factor written on the back of the gray
card.
Figure 4-3, Condor Power Supply (front view)
Bowl Optics
Signal