2-2
PCS2 Service Manual
1. Remove (3) Phillips head screws from the bottom edge of the rear
panel.
2. Pull bottom edge 6" away from the cabinet, then pull the rear panel
down and away from the upper lip of the cabinet.
3. Unplug the fans at P407 on the centrifuge distribution board.
4. Disconnect the cuff pneumatic line at the quick disconnect,
approximately 12 inches from the rear panel.
5. Remove the ground wire from the ground lug on the rear panel.
Fan Assembly
1. Remove the rear panel.
2. Remove (2) nuts for each fan to be removed.
Note:
All four fans should be replaced at the same time, even if only one
is failing.
Cover Stops
1. Remove the rear panel.
2. Remove (2) Phillips head screws for each cover stop to be removed.
Cuff Connector
1. Remove the rear panel.
2. Remove connector nut.
3. Push connector out of panel from rear.
1. Tilt the cabinet back slightly, and remove the (3) Phillips head screws
from the bottom edge of the front panel.
2. Pull the front panel away from the cabinet, then pull the panel down
and away from the upper lip of the cabinet.
3. Disconnect the cuff pneumatic line at the quick disconnect.
4. Disconnect the ribbon cable P503/703 at P703 on the mother
backplane card.
5. Remove the ground wire from the ground lug on the front panel.
Rear Panel
Front Panel