6-6
PCS2 Service Manual
CODE
TEXT MESSAGE
DESCRIPTION
SUGGESTED RESOLUTION
10
11
11
11
12
DPM FAILED 80%
ANALOG TEST
SPM FAILED 20%
ANALOG TEST
SPM FAILED 50%
ANALOG TEST
SPM FAILED 80%
ANALOG TEST
DONOR VALVE
POSITION FAULT
PLASMA VALVE
POSITION FAULT
SALINE VALVE
POSITION FAULT
DPM channel of A/D
converter failed 80%
analog voltage load
test.
SPM channel of A/D
converter failed 20%
analog voltage load
test.
SPM channel of A/D
converter failed 50%
analog voltage load
test.
SPM channel of A/D
converter failed 80%
analog voltage load
test.
Donor valve position
switch indicates that
valve did not
transition.
Plasma valve position
switch indicates that
valve did not
transition.
Saline valve position
switch indicates that
valve did not
transition.
1. Replace processor card. Using Utilities, calibrate all the
components under A/D, centrifuge, safety system, and
configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.
2. Replace weigher load cell and calibrate in A/D in
Utilities.
1. Replace processor card. Using Utilities, calibrate all the
components under A/D, centrifuge, safety system, and
configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.
1. Replace processor card. Using Utilities, calibrate all the
components under A/D, centrifuge, safety system, and
configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.
1. Replace processor card. Using Utilities, calibrate all the
components under A/D, centrifuge, safety system, and
configure the language. Reset protocol parameters
(MODIFY key), if parameters were different from
default values.
1. Use the valve test in Diagnostics to manually transition
valves and monitor the valve and switch response.
2. Swap valve connections at top deck distribution card to
isolate problem to valve or circuit.
3. If valve does not move, check, repair or replace all
connections associated with the appropriate valve drive
circuit.
4. Replace valve, driver card, or processor card. Using
Utilities, calibrate all the components under A/D,
centrifuge, safety system, and configure the language.
Reset protocol parameters (MODIFY key), if parameters
were different from default values.
5. If valve does move but is not recognized, check, repair,
or replace all connections associated with the appropriate
valve position switch or replace the valve.
6. Replace driver or processor card. Using Utilities,
calibrate all the components under A/D, centrifuge,
safety system, and configure the language. Reset
protocol parameters (MODIFY key), if parameters were
different from default values.