A safety system generated error message does not have a text description.
Instead, five safety system registers report information in hexadecimal
format.
Safety system generated error messages resemble this example, where
each XXXXXX represents a hexadecimal code:
The CPU generated error messages are listed on the following pages in
numerical order for those that have associated error codes. The remaining
status and error messages are listed alphabetically. The messages are
accompanied by a brief description of the message, and then followed by
resolution steps to alleviate the problem.
Chapter 6, Troubleshooting
6-3
CPU Error
Messages
SAFETY SYSTEM FAULT DETECTION
COMMAND: XXXXXX
CHANNEL 1
CHANNEL 2
FAULT:
XXXXXX
XXXXXX
CRITICAL:
XXXXXX
XXXXXX
PLEASE RECORD THESE VALUES
Press NO to Mute Safety Alarm
Press YES to Resume Procedure