background image

Technical 
Publication

0366

GE Medical Systems - Kretztechnik

Voluson® 730
Service Manual

Direction 105844
Revision 1

Copyright

©

 2002; 2003 by General Electric Co.

Operating Documentation

GE Medical Systems
Kretz Ultrasound

Содержание Voluson 730

Страница 1: ...al Publication 0366 GE Medical Systems Kretztechnik Voluson 730 Service Manual Direction 105844 Revision 1 Copyright 2002 2003 by General Electric Co Operating Documentation GE Medical Systems Kretz Ultrasound ...

Страница 2: ......

Страница 3: ...LANGUE QUE L ANGLAIS C EST AU CLIENT QU IL INCOMBE DE LE FAIRE TRADUIRE NE PAS TENTER D INTERVENTION SUR LES ÉQUIPEMENTS TANT QUE LE MANUEL SERVICE N A PAS ÉTÉ CONSULTÉ ET COMPRIS LE NON RESPECT DE CET AVERTISSEMENT PEUT ENTRAÎNER CHEZ LE TECHNICIEN L OPÉRATEUR OU LE PATIENT DES BLESSURES DUES À DES DANGERS ÉLECTRIQUES MÉCANIQUES OU AUTRES DIESES KUNDENDIENST HANDBUCH EXISTIERT NUR IN ENGLISCHER S...

Страница 4: ... ESTES MANUAIS NOUTRO IDIOMA É DA RESPONSABILIDADE DO CLIENTE FORNECER OS SERVIÇOS DE TRADUÇÃO NÃO TENTE REPARAR O EQUIPAMENTO SEM TER CONSULTADO E COMPREENDIDO ESTE MANUAL DE ASSISTÊNCIA TÉCNICA O NÃO CUMPRIMENTO DESTE AVISO PODE POR EM PERIGO A SEGURANÇA DO TÉCNICO OPERADOR OU PACIENTE DEVIDO A CHOQUES ELÉTRICOS MECÂNICOS OU OUTROS IL PRESENTE MANUALE DI MANUTENZIONE È DISPONIBILE SOLTANTO IN IN...

Страница 5: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL iii ...

Страница 6: ...ENT FOR USA ONLY All electrical Installations that are preliminary to positioning of the equipment at the site prepared for the equipment shall be performed by licensed electrical contractors Other connections between pieces of electrical equipment calibrations and testing shall be performed by qualified GE Medical Systems personnel In performing all electrical work on these products GEwill use it...

Страница 7: ...pages 3 11 to 3 27 pages 3 28 to 3 34 pages 3 35 to 3 38 1 0 1 0 1 Chapter 8 Replacement Procedures pages 8 1 to 8 8 1 Rev History LOEP pages v to vi 1 Chapter 4 Functional Checks pages 4 1 to 4 22 pages 4 23 to 4 28 pages 4 29 to 4 30 pages 4 31 to 4 32 pages 4 33 to 4 36 0 1 0 1 0 Chapter 9 Replacement Parts pages 9 1 to 9 3 pages 9 4 to 9 29 pages 9 30 to 9 34 0 1 0 Table of Contents pages vii ...

Страница 8: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL vi ...

Страница 9: ... 1 2 Important Conventions 1 3 Conventions Used in Book 1 3 Standard Hazard Icons 1 4 Product Icons 1 5 Safety Considerations 1 7 Introduction 1 7 Human Safety 1 7 Mechanical Safety 1 7 Electrical Safety 1 8 Labels Locations 1 8 Dangerous Procedure Warnings 1 9 Lockout Tagout Requirements For USA Only 1 9 Returning Shipping Probes and Repair Parts 1 9 EMC EMI and ESD 1 10 Electromagnetic Compatibi...

Страница 10: ...ts 2 2 Voluson 730 Power Requirements 2 2 Inrush Current 2 2 Site Circuit Breaker 2 3 Site Power Outlets 2 3 Unit Power Plug 2 3 EMI Limitations 2 4 Scan Probe Environmental Requirements 2 5 Time and Manpower Requirements 2 5 Facility Needs 2 6 Purchaser Responsibilities 2 6 Required Features 2 6 Desirable Features 2 7 Minimal Floor Plan Suggestion 2 8 Networking Pre installation Requirements 2 9 ...

Страница 11: ...tion 3 7 Completing the Installation 3 8 System Specifications 3 8 Physical Dimensions of Voluson 730 3 8 Weight without Monitor and Peripherals 3 8 Acoustic Noise Output 3 8 Electrical Specifications 3 8 Power On Boot Up 3 9 Scanner Power On 3 9 Back end Processor Boot Up 3 9 Power Off Shutdown 3 10 Back end Processor Power Down 3 10 Scanner Shutdown 3 10 Transducer Connection 3 11 Printer Instal...

Страница 12: ...tion 3 28 System Configuration 3 28 How to enter Date and Time 3 29 How to enter Hospital Name 3 29 How to change Language 3 29 On Board Optional Peripherals 3 30 External I O Connection Panel GES 3 31 External I O Pin Outs 3 32 Video Specification 3 34 Available Probes 3 35 Software Option Configuration 3 35 Connectivity Installation Worksheet 3 36 Network IP Address Configuration 3 37 Paperwork ...

Страница 13: ... Volume Mode Functions 4 14 Functions after the 3D Acquisition 4 15 Functions in the Tool Menu 4 16 Using Cine Mode 4 17 Activating Cine 4 17 Cine Split Function Multiple Format 4 17 Activating 2D Auto Cine 4 17 Spectral Doppler or M Cine Loop 4 17 Activating 3D Rotation Cine 4 17 Activating 4D Cine 4 17 Activating Live 3D Cine 4 17 Basic Measurements 4 18 Distance and Tissue DepthMeasurements B a...

Страница 14: ...4 23 Using the MOD Magneto Optical Drive 4 24 Backup and Restore Database Preset Configurations and Images 4 25 Formatting Media 4 25 Save System Presets and Configurations Application Settings 4 27 Load System Presets and Configurations Application Settings 4 28 Archiving Images 4 30 ECG Check Out 4 31 Software Configuration Checks 4 31 Peripheral Checks 4 32 Mechanical Function Checks 4 33 Rotat...

Страница 15: ...cription 5 10 B Mode 5 10 M Mode 5 11 D Mode Pulsed Wave and Continuous Wave Doppler 5 12 CFM Mode Color Flow Mode 5 12 3D Mode 5 13 Real Time 4D Mode 5 13 External Video Mode Video display from VCR 5 14 Description of Software Options 5 15 Real Time 4D 5 15 DICOM 5 15 Real Time 4D Biopsy 5 15 VOCAL Virtual Organ Computer aided Analysis 5 15 Tissue Harmonic Imaging THI 5 16 Sonoview Image Manageme...

Страница 16: ...m CPS 5 25 Back End Board Descriptions 5 26 SBC Single Board Computer 5 26 CPS DMA Controller Card 5 26 CKV Video Card 5 26 Hard Disk Drive 5 26 CPE Back Panel I O Card 5 26 CPM PC Motherboard card 5 26 CPP Power Supply Secondary Board Motor Power stage 5 27 Internal I O 5 28 Top Console 5 31 Monitor 5 32 External I O 5 33 Peripherals 5 34 General Information 5 34 ECG preamplifier MAN6 optional 5 ...

Страница 17: ...GW GEU Monitor 5 39 Air Flow Control 5 40 Air Flow Distribution 5 40 Air Flow Distribution Overview 5 40 Service Page 5 41 Introduction 5 41 Access Security 5 41 Service Login 5 41 Auto Tester File 5 42 Service Viewer 5 43 Export System Status 5 43 Export Event Log 5 43 Delete all Patients 5 43 Maintenance Report 5 43 Export System Data 5 44 Update 5 44 EUM 5 44 NLS 5 44 Printer 5 44 ...

Страница 18: ... 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xvi CHAPTER 6 Service Adjustments Overview 6 1 Purpose of Chapter 6 6 1 Regulatory 6 1 Monitor Adjustment 6 2 Control Console Transport Lock 6 4 Control Console 6 4 Transport Lock 6 4 Trackball Adjustment 6 5 ...

Страница 19: ... 4 Export System Data 7 4 Export System Status 7 4 How to use the Auto Tester program 7 5 Minimum Cable Configuration to Boot 7 8 Minimum Cable Configuration to Scan 7 9 Troubleshooting Trees and Instructions 7 10 System does not Power On Boot Up 7 10 Noise in Image 7 11 Trackball 7 12 System does not Power Off Shutdown 7 13 Monitor Troubleshooting 7 14 Unable to Record to VCR 7 15 Printer Trouble...

Страница 20: ...nd Repair Parts 8 1 Backup Loading Procedure 8 2 Loading Procedure 8 2 EUM Electronic User Manual Upgrade Procedure 8 3 EUM Upgrade Procedure up to Software Version 1 06F 8 3 EUM Upgrade Procedure from Software Version 1 06G 8 5 EUM Removal Procedure 8 5 EUM Installation Procedure 8 7 Replacement of the Trackball top fixation ring 8 8 Manpower 8 8 Trackball top fixation ring Replacement Procedure ...

Страница 21: ...ps 9 3 Housing GW and additional Console Hardware 9 4 User Interface GEU Monitor 9 6 Disk Drives GEM 9 9 Main Power Module CPN 9 10 Main Board Module GEF 9 11 FrontEnd US Part 9 12 BackEnd Processor PC Part 9 16 Options and Upgrades 9 18 Miscellaneous Cables 9 19 Optional Peripherals and Accessories 9 25 Probes 9 29 2D Probes 9 29 Real Time 4D Volume Probes 9 31 CW Pencil Probes 9 32 Live 3D Probe...

Страница 22: ...Maintenance 10 6 Preliminary Checks 10 6 Functional Checks See Also Chapter 4 10 7 System Checks 10 7 Peripheral Option Checks 10 8 Input Power 10 8 Mains Cable Inspection 10 8 Cleaning 10 8 General Cleaning 10 8 Physical Inspection 10 9 Optional Diagnostic Checks 10 10 Probe Maintenance 10 10 Probe Related Checks 10 10 Basic Probe Care 10 10 Basic Probe Cleaning 10 10 Using a Phantom 10 11 Electr...

Страница 23: ... 17 Definition 10 17 Generic Procedure 10 18 Isolated Patient Lead Source Leakage Lead to Lead 10 18 Isolated Patient Lead Sink Leakage Isolation Test 10 19 Data Sheet for ECG Leakage Current 10 20 Probe Leakage Current Test 10 21 Definition 10 21 Generic Procedure 10 21 No Meter Probe Adapter Procedure 10 22 Data Sheet for Transducer Source Leakage Current 10 23 When There s Too Much Leakage Curr...

Страница 24: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL xxii ...

Страница 25: ...nctional Checks Contains functional checks that are recommended as part of the installation or as required during servicing and periodic maintenance 5 Chapter 5 Components and Functions Theory Contains block diagrams and functional explanations of the electronics 6 Chapter 6 Service Adjustments Contains instructions on how to make available adjustments to the Voluson 730 7 Chapter 7 Diagnostics Tr...

Страница 26: ...tc Hospital s Service Personnel Contractors Some parts of Chapter 2 Pre Installation 1 1 4 Voluson 730 Models Covered by this Manual 1 1 5 Purpose of Operator Manual s The Operator Manual s should be fully read and understood before operating the Voluson 730 and also kept near the unit for quick reference Table 1 2 Voluson 730 Model Designations GE Part Number Kretz Description H46601A 154585 Volu...

Страница 27: ...wn or potential hazards are labeled in one of following ways NOTE Notes provide important information about an item or a procedure Information contained in a NOTE can often save you time or effort DANGER DANGER IS USED TO INDICATE THE PRESENCE OF A HAZARD THAT WILL CAUSE SEVERE PERSONAL INJURY OR DEATH IF THE INSTRUCTIONS ARE IGNORED WARNING WARNING WARNING IS USED TO INDICATE THE PRESENCE OF A HA...

Страница 28: ...chapter In addition to text several different graphical icons symbols may be used to make you aware of specific types of hazards that could cause harm Other hazard icons make you aware of specific procedures that should be followed Table 1 3 Standard Hazard Icons ELECTRICAL MECHANICAL RADIATION LASER HEAT PINCH Table 1 4 Standard Icons Indicating a Special Procedure Be Used AVOID STATIC ELECTRICIT...

Страница 29: ...IPX 7 Indicates the degree of protection provided by the enclosure per IEC 529 IPX 1 and IPX 7 indicates drip proof Footswitch Probes Equipment Type BF man in the box symbol IEC 878 02 03 indicates B Type equipment having even more electrical isolation than standard Type B equipment because it is intended for intimate patient contact Probe connectors Front side of the ECG preamplifier MAN Rear of ...

Страница 30: ...to On Off Standby switch left below the Control panel Mains ON Indicates the power on position of the mains power switch Rear of system at mains switch F1 ON switch of the isolation transformer for auxiliary devices Rear of system at the switch for auxiliary devices F2 OFF switch of the isolation transformer for auxiliary devices Rear of system at the switch for auxiliary devices F2 Protective Ear...

Страница 31: ...MOVED ALONG ANY INCLINE USE EXTREME CAUTION SINCE IT MAY BECOME UNSTABLE AND TIP OVER WARNING WARNING ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN EASILY BE DAMAGED BY IMPROPER HANDLING USE CARE WHEN HANDLING AND PROTECT FROM DAMAGE WHEN NOT IN USE DO NOT USE A DAMAGED OR DEFECTIVE PROBE FAILURE TO FOLLOW THESE PRECAUTIONS CAN RESULT IN SERIOUS INJURY AND EQUIPMENT DAMAGE WA...

Страница 32: ...s equipped with a three conductor AC power cable This must be plugged into an approved electrical outlet with safety ground If an extension cord is used with the system make sure that the total current rating of the system does not exceed the extension cord rating The power outlet used for this equipment should not be shared with other types of equipment Both the system power cable and the power c...

Страница 33: ...t with visible body fluids be taken or shipped from a clinic or site for example body coils or and ultrasound probe The purpose of the regulation is to protect employees in the transportation industry as well as the people who will receive or open this package The US Department of Transportation DOT has ruled that items what were saturated and or dripping with human blood that are now caked with d...

Страница 34: ...ludes possible effects from electrical fields magnetic fields electrostatic discharge and disturbances in the electrical power supply For applicable standards refer to Chapter 2 in the Basic User Manual 1 4 2 CE Compliance The Voluson 730 unit conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge radiated and conducted RF fields magnetic fie...

Страница 35: ...e Location Phone Number USA Canada GE Medical Systems Ultrasound Service Engineering 4855 W Electric Avenue Milwaukee WI 53219 Customer Answer Center 1 800 437 1171 1 800 682 5327 1 262 524 5698 Fax 1 414 647 4125 Latin America GE Medical Systems Ultrasound Service Engineering 4855 W Electric Avenue Milwaukee WI 53219 Customer Answer Center 1 262 524 5300 1 262 524 5698 Fax 1 414 647 4125 Europe G...

Страница 36: ...VISION 1 VOLUSON 730 SERVICE MANUAL 1 12 Section 1 5 Customer Assistance 1 5 2 System Manufacturer Table 1 7 System Manufacturer Manufacturer Telephone FAX GE Medical Systems Kretztechnik GmbH Co OHG Tiefenbach 15 A 4871 Zipf Austria 43 7682 3800 0 43 7682 3800 47 ...

Страница 37: ...w 2 1 1 Purpose of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a Voluson 730 Included are descriptions of the facility and electrical needs to be met by the purchaser of the unit Table 2 1 Contents in Chapter 2 Section Description Page Number 2 1 Overview 2 1 2 2 General Console Requirements 2 2 2 3 Facility Needs 2 6 ...

Страница 38: ...edicated power and ground for the proper operation of its Ultrasound equipment This dedicated power shall originate at the last distribution panel before the system The dedicated line shall consist of one phase a neutral not shared with any other circuit and a full size ground wire from the distribution panel to the Ultrasound outlet Please note that image artifacts can occur if at any time within...

Страница 39: ...dical and test equipment needed to support this unit must also be present within 1 m 3 2 ft of the unit Electrical installation must meet all current local state and national electrical codes 2 2 2 5 Unit Power Plug If the unit arrives without a power plug or with the wrong plug you must contact your GEdealer or the installation engineer must supply what is locally required CAUTION POWER OUTAGE MA...

Страница 40: ...t Poor grounding is the most likely reason a unit will have noisy images Check grounding of the power cord and power outlet Replace all screws RF gaskets covers cores After you finish repairing or updating the system replace all covers and tighten all screws Any cable with an external connection requires a magnet wrap at each end Install the shield over the front of card cage Loose or missing cove...

Страница 41: ...takes time Begin Pre installation checks as soon as possible If possible allow six weeks before delivery for enough time to make necessary changes NOTICE SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF 10 TO 50 degrees C WHEN EXPOSED TO LARGE TEMPERATURE VARIATIONS THE PRODUCT SHOULD BE KEPT IN ROOM TEMPERATURE FOR 10 HOURS BEFORE USE CAUTION Have two people available to del...

Страница 42: ...listed or customer provided product or to place an approved product further from the system than the interface kit allows presents challenges to the installation team To avoid delays during installation such variances should be made known to the individuals or group performing the installation at the earliest possible date preferably prior to the purchase The ultrasound suite must be clean prior t...

Страница 43: ...cm 3 ft wide Circuit breaker for dedicated power outlet is easily accessible Sink with hot and cold water Receptacle for bio hazardous waste like used probe sheaths Emergency oxygen supply Storage for linens and equipment Nearby waiting room lavatory and dressing room Dual level lighting bright and dim Lockable cabinet ordered by GE for its software and proprietary manuals ...

Страница 44: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL 2 8 Section 2 3 Facility Needs 2 3 4 Minimal Floor Plan Suggestion Figure 2 1 Minimal Floor Plan ...

Страница 45: ...an added benefit transferring images in this manner frees up the on board monitor and peripherals enabling viewing to be done while scanning continues With DICOM images can be archived stored and retrieved faster easier and at a lower cost 2 3 5 2 DICOM Option Pre installation Requirements To configure the Voluson 730 to work with other network connections the site s network administrator must pro...

Страница 46: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL 2 10 Section 2 3 Facility Needs This page was intentionally left blank ...

Страница 47: ...nts in Chapter 3 Section Description Page Number 3 1 Overview 3 1 3 2 Receiving and Unpacking the Equipment 3 4 3 3 Preparing for Installation 3 6 3 4 Completing the Installation 3 8 3 5 Printer Installation 3 12 3 6 System Configuration 3 28 3 7 Connectivity Installation Worksheet 3 36 3 8 Network IP Address Configuration 3 37 3 9 Paperwork 3 38 Table 3 2 Average Installation Time Description Ave...

Страница 48: ...xtended position the console could tip over CAUTION Equipment damage possibility Turning the system on without acclimation after arriving at site may cause the system to be damaged Table 3 3 Acclimation Time C 60 55 50 45 40 35 30 25 20 15 10 5 0 5 10 15 20 25 30 35 40 F 140 131 122 113 104 96 86 77 68 59 50 41 32 23 14 5 4 13 22 31 40 hrs 8 6 4 2 0 0 0 0 0 0 0 2 4 6 8 10 12 14 16 18 20 Figure 3 1...

Страница 49: ...power outlet Do not use a three to two prong adapter This defeats safety grounding CAUTION Do NOT wear the ESD wrist strap when you work on live circuits and more than 30 V peak is present CAUTION Do not use a20 Amp to 15 Amp adapter on the 120 Vac unit s power cord This unit requires a dedicated 16 A circuit CAUTION Do not operate this unit unless all board covers and frame panels are securely in...

Страница 50: ...luson 730 Attempts to move the unit considerable distances or on an incline by one person could result in injury or damage or both Figure 3 3 Unpacking the system Table 3 4 Unpacking Procedure Step Task 1 Loosen the screws and remove top panel from crater Unpack the devices such a way that packaging can be reused The device must only be transported in the original packaging Check whether delivery ...

Страница 51: ... are needed in the next step due to the weight of the equipment 4 Disengage the brakes and slowly move unit down the ramp Note Packing crate and material should be stored for future use 5 Remove Monitor from its box and mount it on the system Table 3 4 Unpacking Procedure Step Task screws screws ramp packing Left and Right panels are shown removed for clarity Brakes Brakes Ramp Monitor Mounting Br...

Страница 52: ...e power switch on the rear side of the monitor NOTE The switch of the monitor has to be in ON position before starting the system Leave monitor power switch always ON 3 3 1 2 Connection of Peripherals 1 Remove rear cover plate see Figure 3 5 Carefully remove plastic caps using a slotted screwdriver or the blade of a knife Loosen screws and washers 2 When the cover is loose on top pull the rear cov...

Страница 53: ...rdered or damaged 3 3 3 Physical Inspection 3 3 3 1 System Voltage Settings Verify that the scanner is set to the correct voltage The Voltage settings for the Voluson 730 Scanner is found on a metal plate on the rear of the system 3 3 4 EMI Protection This unit has been designed to minimize the effects of Electo Magnetic Interference EMI Many of the covers shields and screws are provided primarily...

Страница 54: ...T1 ST2 ST3 ST4 ST5 for accessories All mains outlets are co switched by the unit s mains switch via built in isolation transformer Output voltage for ST1 ST5 115V or 230V Table 3 5 Physical Dimensions of Voluson 730 Height Width Depth 145 cm 57 1 inches 68 cm 26 8 inches 100 cm 39 4 inches Table 3 6 Weight of Voluson 730 With Monitor and Without Other Peripherals Model Weight kg Weight lbs Voluson...

Страница 55: ...o initiate the scanner The company logo appears on the Touch screen during boot up Depending on the BIOS Version no status messages are displayed during this process Boot up time is about 3 minutes NOTE The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the Stand By switch The switch of printers has to be in ON position before starting the system So the au...

Страница 56: ... the system for peripheral auxiliary equipment are commonly switched with the Standby switch So the auxiliary equipment need not to be switched ON OFF separately 3 Disconnect the Mains Power Cable if necessary For example Repairing the system 4 Press once on the brakes to block the front wheels brakes on front wheels under the foot rest 5 Prior to disconnect a probe freeze the image 6 Open the rig...

Страница 57: ...er twist lock lever is at horizontal position 2 Insert the transducer connector on the receptacle guide pin until it touches the receptacle mating surface 3 Twist the transducer twist lock lever to vertical position to lock it in place Twist the lever to the horizontal position to disconnect the transducer 4 Open the right hand side door lay the cable into the intended cable holder and close the d...

Страница 58: ...ait till the system has booted 2 Prepare the color printer as illustrated at HP quick start poster Windows quick start Attach Automatic Two Side Printing Module and load white paper Plug in power cord Do NOT turn ON the printer Turn on the printer and install print cartridges after closing the top cover a calibration page is printed automatically 3 5 1 1 Install the HP printer software driver Perf...

Страница 59: ...f no drivers can be found on the hard disk C for HP995c use the HP Installation disk and browse for the file HPF995K INF Path F DEU DRIVERS WIN2K_XP 6 Confirm the correct path and click NEXT to install the driver All necessary files are copied 7 Confirm the installation by clicking FINISH to close the Hardware Wizard 8 Close all open windows close the System Setup with SAVE EXIT and restart the sy...

Страница 60: ...keys and for opening a context menu press the upper trackball key 3 Turn ON the printer then switch ON the power of the system and wait till the system has booted NOTE The power switch of the printer has to be in ON position before starting the system 3 5 2 1 Install the UP D21MD printer software driver Perform the following steps if this printer was never installed on the Voluson 730 4 Connect th...

Страница 61: ... to install the driver All necessary files are copied 8 The Message Digital Signature Not Found appears Click YES 9 Confirm the installation by clicking FINISH to close the Hardware Wizard 10 Close all open windows close the System Setup with SAVE EXIT and restart the system turn off and on the system Figure 3 16 Search for Network path Figure 3 17 Confirm correct path and finish the Installation ...

Страница 62: ...ctions can be performed by using the trackball for moving the cursor Normal select Click by pressing the left or right trackball keys and for opening a context menu press the upper trackball key 7 Turn ON the printer 8 Switch ON the power of the system and wait till the system has booted NOTE The power switch of the printer has to be in ON position before starting the system Leave the power switch...

Страница 63: ...Panel press UTILITIES 2 In the Utilities menu touch SYSTEM SETUP to invoke the setup desktop on the screen 3 Select the SERVICE page The password window appears automatically 4 Enter the password SHE and click the ACCEPT button to display the Service Tools window 5 Click on the PRINTER button Figure 3 18 System Setup Service page Figure 3 19 Service Tools window Printer button ...

Страница 64: ...lly cont d 6 Click the ADD PRINTER button A warning message appears Please read this message carefully and click YES if you have skills to do this Figure 3 20 Printer Installation Properties NOTICE At older Software versions first the warning message appears After confirming with the YES button click the ADD PRINTER icon on the displayed screen Figure 3 21 Add Printer ...

Страница 65: ...rinter and then click NEXT 9 Select the corresponding Printer Port e g Figure 3 23 Sony UP D21MD USB002 and click NEXT Figure 3 22 Add Printer Wizard CAUTION If you have to install the SCSI Printer UP D2600S or the SCSI Printer CP770DW for the first time the correct port may not be listed In this case select any port and finish the installation procedure After restart of the system adjust the port...

Страница 66: ...rinterDriver HPDeskjet990c for Line Printer HP 995c C Utilities PrinterDriver HPDeskjet995c for Color Printer UP D21MD C Utilities PrinterDriver SonyUP D21MD for SCSI Printer UP D2600S C Utilities PrinterDriver SonyUP D2600S for SCSI Printer CP770DW C Utilities PrinterDriver MitsubishiCP770DW NOTE If no drivers can be found on the hard disk C for HP995c use the HP Installation disk and browse for ...

Страница 67: ...confirm with OK 13 Select the manufacturer and model of your printer and confirm with the NEXT button 14 Assign a name decide if the printer should be used as default printer and confirm with NEXT see Figure 3 27 15 Select Do not share this printer and confirm the Printer Sharing window Figure 3 27 by clicking NEXT Figure 3 26 verify selected Printer Driver path Figure 3 27 Assign name and select ...

Страница 68: ...rd window appears on the screen 17 Complete the manual Printer Installation with the FINISH button 18 Close all open windows close the System Setup with SAVE EXIT and restart the system turn off and on the system Figure 3 28 Complete manual Printer Installation NOTICE Ver After boot up of the system verify the correct settings in the printer Properties see Section 3 5 5 Adjustment of Printer Setti...

Страница 69: ... adjust the HP 990cxi HP 995c Printer see Section 3 5 5 1 HP 990cxi HP 995c Printer Settings To adjust the UP D21MD Printer see Section 3 5 5 2 UP D21MD Printer Settings To adjust the CP770DW Printer see Section 3 5 5 3 CP770DW Printer Settings Figure 3 29 select the desired printer NOTICE At older Software versions first the warning message appears After confirming with the YES button select the ...

Страница 70: ...ATURES page for adjusting Two Side Printing if desired Orientation must be set to Portrait See right Figure 3 31 5 For saving the adjusted printer settings click APPLY and then OK 6 Select the PORTS page and selct verify the correct USB printer port 7 For saving the adjusted printer settings click APPLY and then OK Finally close the Printers window with the close button and exit System Setup with ...

Страница 71: ... High Speed 3 Select the GRAPHICS page select Gamma Select from the pop up menu choose Gamma 1 4 Select Color Correction from the pop up menu set Printer Hardware Color Correction 5 Select Lightness from the pop up menu Sharpness 7 or 8 Dark 0 Gamma 12 Light 8 6 For saving the adjusted printer settings click APPLY and then OK Finally close the Printers window with the close button and exit System ...

Страница 72: ...ttings 2 Select the PORTS page and select verify the correct printer port 3 Select the GENERAL page and click the PRINTING PREFERENCES button 4 In the displayed PAPER page select Paper Size S and Orientation Portrait 5 Select the COLOR ADJUST page and click the GAMMA ADJ button 6 Hook the Adjust field and move the line with the trackball till it shows Input 65 Output 100 Figure 3 34 Ports page Pap...

Страница 73: ... key Remote PRINT B key and Report Printer to desired Printer 1 On the Touch Panel touch UTILITIES 2 In the Utilities menu touch SYSTEM SETUP to invoke the setup desktop on the screen 3 Select the PERIPHERALS page Remote Print A Select the desired Printer for the remote control PRINT A key Remote Print B Select the desired Printer for the remote control PRINT B key NOTE Optionally the Remote Contr...

Страница 74: ...options are installed in the system Service enter the password to get access to the Service Tools functions Backup Backup Save Backup Load DICOM to set up all DICOM target nodes image servers System Info shows which Software Hardware version is installed in the system To invoke the System Setup procedure 1 Touch the UTILITIES key on the Touch Panel 2 Touch the SYSTEM SETUP key in the Utilities men...

Страница 75: ... enter Hospital Name Select the GENERAL page in the System Setup see Figure 3 37 1 Select the text box to enter a new Clinic Name with the keyboard 2 Click SAVE EXIT to save Settings and exit System Setup The clinic name will be copied into the Hospital ID in the information header 3 6 1 3 How to change Language Select the GENERAL page in the System Setup see Figure 3 37 1 Select the desired langu...

Страница 76: ... A 50 60 Hz 2 240 VAC 10 4 20 A 50 60 Hz 2 Table 3 9 Approved Peripherals Device Manufacturer Model Video Signal B W Video Printer SONY UP 895MD NTSC PAL B W Video Printer Mitsubishi P91E NTSC PAL Color Video Printer SONY UP D2600S SCSI Port Color Video Printer SONY UP D21MD USB Port Color Video Printer Mitsubishi CP770DW SCSI Port Video Cassette Recorder SONY SVO 9500MD SVO 9500 MDP NTSC PAL Line...

Страница 77: ... External I O Panel Connectors Table 3 10 External I O Connector Description Item Connector Name Table Number Description 1 VGA OUTPUT Table 3 13 print out VGA signal with monitor printer 2 R G B H V SYNC Table 3 16 outputs for color video printer monitor 3 NETWORK Table 3 15 DICOM input output twisted pair RJ 45 10 100 megabit s 4 USB 1 Table 3 14 USB port 5 USB 2 Table 3 14 USB port 6 AUDIO LEFT...

Страница 78: ...AL Table 3 12 S VHS OUT Video Connector 4 Pin Pin No Output Signal Description 1 SVIDEO OUT YG Y Luma GND 2 SVIDEO OUT CG C Chroma GND 3 SVIDEO OUT Y Y Luma SIGNAL 4 SVIDEO OUT C C Chroma SIGNAL Table 3 13 VGA Output Connector Sub D 15 Pin Pin No Output Signal Description 1 VGA OUT1 R Red 2 VGA OUT1 G Green 3 VGA OUT1 B Blue 4 9 11 12 15 N C N C 5 6 7 8 10 GND GND 13 VGA OUT1 HS H Sync 14 VGA OUT1...

Страница 79: ...2 ETHER TD Ethernet RD 3 ETHER RD Ethernet TD 6 ETHER RD Ethernet TD Others NC Non connection Table 3 16 R G B SYNC L Audio R IN OUT VHS IN OUT Connectors Pin No Signal Description 1 Center Pin Signal 2 Koax Pin Signal GND Table 3 17 Footswitch Connector located at Power Supply Module CPN rear side Pin No Input Signal Description 1 Signal GND 2 right switch normally open 3 left switch normally ope...

Страница 80: ...µs 0 616 0 512 Horizontal pos pos Vertical Rate Hz 75 00 90 00 Vertical Total cycle time ms 13 30 11 10 V Sync Width ms 0 170 0 140 V Back Porch ms 0 276 0 332 V Active Video Time ms 12 68 10 48 V Front Porch ms 0 174 0 144 Dot Clock MHz 49 54 58 39 Table 3 19 Video Specifications for Composite B W S Video and RGB Connectors S Video Output set to Timing Parameter PAL 50Hz NTSC 60Hz Visible Resolut...

Страница 81: ... Software Options refer to the Voluson 730 Basic User Manual Chapter 17 3 4 System Setup Options For information on configuring DICOM Connectivity refer to the Voluson 730 Basic User Manual Chapter 17 3 8 System Setup DICOM For information on configuring General Measurement Setup refer to the Voluson 730 Basic User Manual Chapter 18 3 1 For information on configuring Obstetric Measurement Setup re...

Страница 82: ...ion Worksheet Site System Information Services Destination Devices TCP IP Settings Device Type Manufacturer Name IP Address Port AE Title 1 2 3 4 5 6 7 8 9 10 11 12 Name AE Title Dept Serial Type Floor Room REV Comments CONTACT INFORMATION Name Title Phone E Mail Address Site IP Settings IP Address Subnet Mask Default Gateway Remote Archive Setup Remote Archive IP Remote Archive Name ...

Страница 83: ...ties menu 2 Select SYSTEM SETUP and open the DICOM page on the System Setup desktop screen 3 Click the NETWORK CONFIG button read the message and afterwards confirm with YES 4 The Internet Protocol TCP IP Properties dialog page appears To specify a DICOM Address follow the instructions of described in the Basic User Manual Chapter 17 3 8 of the Voluson 730 Figure 3 39 Network Configuration Figure ...

Страница 84: ...all relevant safety and user information is available during the operation and service of the complete system 3 9 1 Product Locator Installation NOTE The Product Locator Installation Card shown may not be same as the provided Product Locator card 3 9 2 User Manual s Check that the correct User Manual s for the system and software revision is included with the installation Specific language version...

Страница 85: ...ired Equipment An empty blank MO Disk or CD RW At least one transducer See Probes on page 9 29 for an overview normally you should check all the transducers used on the system Table 4 1 Contents in Chapter 4 Section Description Page Number 4 1 Overview 4 1 4 2 Required Equipment 4 1 4 3 General Procedure 4 2 4 4 Functional Checks 4 7 4 5 Software Configuration Checks 4 31 4 6 Peripheral Checks 4 3...

Страница 86: ...t Breaker at the rear of the system CAUTION SYSTEM REQUIRES ALL COVERS Operate this unit only when all board covers and frame panels are securely in place The covers are required for safe operation good system performance and cooling purposes NOTICE Lockout Tagout Requirements For USA only Follow OSHA Lockout Tagout requirements by ensuring you are in total control of the Power Cable on the system...

Страница 87: ...tes NOTE The mains outlet of the system for peripheral auxiliary equipment are commonly switched with the Standby switch So the auxiliary equipment need not to be switched ON OFF separately 4 3 2 Power Off Shutdown NOTE After turning off a system wait at least ten seconds before turning it on again The system may not be able to boot if power is recycled too quickly 4 3 2 1 Scanner Shutdown 1 Press...

Страница 88: ... 3D 4D Volume Mode key 6 Depth Focus and Power toggle switch controls 7 VCR Remote Control key 8 Print A Trigger key 9 Inter memory key to save to Sonoview or send to DICOM server 10 Freeze Run key 11 Trackball and Trackball keys 12 Annotation and Measurement keys 13 Keyboard and F1 key to invoke the electronic user manual EUM 14 Probe key 15 Patient Data Entry key 16 Network key Sonoview 17 Print...

Страница 89: ...ing window shows all settings for the active application The active one is highlighted 5 Additional functions which are supported by the selected Mode NOTE Different menus are displayed depending on which Touch Panel Menu is selected At the bottom of the Touch Panel there are combination rotary dials push buttons and flip switch controls The functionality of these controls changes depending upon t...

Страница 90: ...splay Tour Table 4 2 Monitor Display Features 1 Patient Name family given prefix 10 Focal zone marker s 2 Hospital Identification 11 Gray scale wedge 3 Date 12 Depth scale markers 4 Time 13 Image Info 5 Depth Frame rate 14 TGC curve 6 Probe Application 15 Image Area 7 Patient Identification ID number 16 Status Bar 8 Logo 17 Measurement results 9 Orientation marker 18 Bodymarks ...

Страница 91: ... Color Doppler Modes CFM Color Flow Mode PD Power Doppler TD Tissue Doppler Volume Modes 3D Real Time 4D Live 3D Real Time 4D Biopsy VCI Volume Contrast Imaging NOTE Some software may be considered standard depending upon system configuration If any Modes or Options are not part of the system configuration the check can be omitted NOTE Different menus are displayed depending on which Touch Panel M...

Страница 92: ... acoustic window but also in oblique directions The advantages of CRI are enhanced contrast resolution with better tissue differentiation and clear organ borders 8 LINEAR TRAPEZ Advantage of the Trapezoid Mode The scan area is very increased in relation to the linear display by steering the ultrasound lines in the border of the probe 9 Image Orientation Use the LEFT RIGHT respectively the UP DOWN ...

Страница 93: ...l to noise ratio 19 DYN CONTR Dynamic Range controls how echo intensities are converted to shades of gray thereby increasing the adjustable range of contrast 20 ENHANCE Edge Enhance brings out subtle tissue differences and boundaries by enhancing the gray scale differences corresponding to the edges of structures Adjustments to M Mode s edge enhancement affects the M Mode only 21 REJECT Selects a ...

Страница 94: ...kball key to activate both Modes B M 3 M Mode Gain Rotate the M MODE key to adjust the sensitivity brightness of the entire M image 4 M Mode Depth Common with B Mode Depth 5 SPEED By touching up or down four different sweep speeds can be selected 6 INVERT Invert of the M Mode image 7 FREQUENCY Common with B Mode Frequency 8 FORMAT For selection of three different rations of display format 9 DYN CO...

Страница 95: ...ime Auto Trace The envelope curve of the Doppler spectrum maximum velocities and the corresponding evaluations are automatically displayed on the monitor 7 INVERT To invert the Doppler spectrum display in relation to the direction of the flow 8 ANGLE The angle cursor can be turned in both directions without stop By pressing the angle knob repeatedly the angle correction switches from 60 to 0 and t...

Страница 96: ...e Power Doppler Gain Rotate the C MODE key to adjust the Tissue Doppler Gain 3 STEERING Beam Steering is only possible with linear probes in CFM and PD Mode 4 2D 2D C PD or TD Changes the Single image display to two simultaneous half images The left frame shows only the 2D Mode image The right frame shows the 2D Mode image with color information 5 INVERT The color of the color wedge inverts around...

Страница 97: ...the display It allows you to enhance an interesting part of the grayscale 16 LINE DEN Determines the line density within the Color Box The lower the line density the larger the line distance and the size of the color pixels 17 SCALE CFM TD The maximum velocities are displayed above and under the color scale in kHz cm s or m s 18 CFM MAP PD TD Provides selectability of the color coding for an optim...

Страница 98: ...ion with PD Mode 2 4D Real time 4D continuous volume acquisition and parallel calculation of 3D rendered images 3 LIVE 3D Live 3D continuous volume acquisition and parallel calculation of 3D rendered images 4 VCI Volume Contrast Imaging improves the contrast resolution and therefore facilitates finding of diffuse lesions in organs 5 4D BIOPSY Real time 4D Biopsy continuous volume acquisition and p...

Страница 99: ...Results 1 M MODE rotary control Rotation about X axis of the reference image PD MODE rotary control Rotation about the Y axis of the reference image PW MODE rotary control Rotation about the Z axis of the reference image 2 C MODE rotary control Movement along Z axis of the reference image TRACKBALL Movement along X and Y axis of the reference image 3 Parts of the orthogonal sections A B and C are ...

Страница 100: ... VCI Figure 4 13 Tool Menu Table 4 9 Functions in the Tool Menu Step Task Expected Results 1 RENDER MODE To select the Render Mode Image Type and Render Algorithm 2 MAGI CUT Ability to electronically manipulate the images and cut way 3D artifacts 3 GRAY CHROMA MAP Depending on individual requirements a harder or softer image can be obtained 4 RENDER DIRECTION To select the desired Render Direction...

Страница 101: ... sequence move the TRACKBALL to display the images one by one 4 4 6 4 Spectral Doppler or M Cine Loop Press FREEZE then roll the TRACKBALL to display the Cine Loop one by one The UPPER TRACKBALL KEY changes from the 2D Cine to the D Loop respectively M Loop NOTE The active Cine is displayed on the monitor screen 2D D M Image or 2D D M Image 4 4 6 5 Activating 3D Rotation Cine 1 After 3D Volume acq...

Страница 102: ...isplays the distance or tissue depth value in the measurement results window Before you complete a measurement To toggle between active calipers press CHANGE upper trackball key To erase results touch the DELETE key on the Touch Panel or press the CLEAR ALL key on the control panel NOTE The CHANGE key alternates the control from one cursor to the other NOTE To exit Basic measurements touch the EXI...

Страница 103: ...t the measurement function without completing the measurement touch EXIT on the Touch Panel or press the MM key again 4 4 7 3 Volume Measurements 1 Press the MM key once 2 Select the appropriate item among 1 DIST 1 ELLIP 3 DIST 1 DIST ELLIP 3 Perform the measurement s using the TRACKBALL and SET right or left trackball key For further details see 4 4 7 1 and 4 4 7 2 4 4 7 3 1 3D Multi Plane Measur...

Страница 104: ...lete the measurement To readjust the traced line press UNDO upper trackball key repeatedly NOTE Depending on the setting in the Measure Setup the envelope curve will be performed with a continuous trace line or by setting points 4 4 7 4 3 Average Velocity Auto Trace 1 Press the MM key once 2 Touch D AUTO TRACE on the Touch Panel It traces the Spectral Doppler image automatically and displays the r...

Страница 105: ...Voluson 730 Basic User Manual Chapter 14 9 Vascular Calculations 4 4 8 5 Report Pages General remarks to perform Calculation measurements By pressing the CALC key on the control panel the Calculation function is switched on Positioning of measurement marks is done with the TRACKBALL Entering and storage of measuring marks is done with SET right or left trackball key To change measuring marks befor...

Страница 106: ...ing key causes loading of the preset The probe activates in the 2D Mode the Touch Panel shows the main menu and the ultrasound image appears on the monitor in Write Mode real time display 4 4 9 3 Deactivating the probe When deactivating the probe the probe is automatically placed in standby mode Read Mode 1 Press the FREEZE key 2 Gently wipe the excess gel from the face of the probe Refer to the U...

Страница 107: ...730 Basic User Manual It talks about several topics Clipboard Sending Exams Printing Exams Images Export Exams Images Backup Exams Restore the Backup Exams DICOM Print Send Sending Images via e mail Browsing and Managing an Exam s stored Image Connectivity and Dataflow Concept and Creation Configuring Connectivity MO and CD RW Formatting Services Destinations Buttons Views Verifying and Pinging a ...

Страница 108: ...o eject a disk from the MOD Manual ejection methods are listed below in preferred order from best 1 to worst 3 a Press the EJECT switch on the MOD while system is ON b Press and hold the EJECT switch while the system is booting c Mechanical ejection Insert the end of a paper clip into the hole next to the EJECT switch while system power is OFF Figure 4 15 Magneto Optical Drive NOTICE Never move th...

Страница 109: ...desired mark the Low Level Format icon 2 Click the START button to start the formatting process 3 A message box appears on the screen Confirm with OK 4 When the formatting has been completed click OK to continue 5 Select the CLOSE key and touch the EXIT key on the Touch Panel to return to the Scan Mode To format the backup media MOD or CD RW press the NETWORK key on the Control panel The Sonoview ...

Страница 110: ...he Touch Panel to return to the Scan Mode NOTE It is possible to adjust the Write Speed of the CD Writer At older software versions also the Erase Mode can be preset in the Settings window 1 Click the SETTINGS icon on the left side of the Sonoview screen 2 Select CD RECORDER from the tool bar 3 Choose the Write Speed and the Erase Mode at systems with older software versions from the corresponding...

Страница 111: ...he customer was using prior to service 1 Insert a formatted MOD Magneto Optical Disk into the drive 2 On the Touch Panel touch UTILITIES 3 On the Utilities Touch Panel touch SYSTEM SETUP to invoke the setup desktop on the screen 4 Select the BACKUP page and click the SAVE button to display the Backup Save window 5 Choose the media e g MO Drive and click the SAVE button 6 Select the NEW FILE button...

Страница 112: ...CAUTION Always backup any Application Settings before the upgrade see Section 4 4 12 2 on page 4 27 The loading procedure overwrites the existing application settings on the local hard drive Make sure to insert the correct MOD or CD RW Additionally you can load the backup from d User Settings Figure 4 23 Backup Load window NOTICE If the selected Backup file doesn t fit a message box appears see Fi...

Страница 113: ...em NOTE Also only parts of a Backup can be loaded into the database to overwrite restore copy etc the database in the system 1 Click the sign to open the content tree 2 Click the ARROW button to copy the selected item into the Load Data field 3 To return selected items from the Load Data field to Backup Data field select the SCISSORS button 4 Click the LOAD button to start loading procedure of the...

Страница 114: ...he TRACKBALL and the right trackball key SET NOTE The number of all exams of the exam list the number of currently selected exams the number of images and the capacity of selected images are displayed automatically at the right and upper corner of the exams list 6 Click the BACKUP button 7 Select on which medium you would like to Backup the files 8 After finishing the backup select whether the sel...

Страница 115: ...desired Language is displayed 4 User Setting Check all the User Settings settings assigned as desired by the customer 5 Peripherals Check assignment of Printer Keys Print A and Print B keys are assigned as desired by the customer 6 Peripherals Check assignment of Foot Switch Foot Switch left and right are assigned as desired by the customer 7 Peripherals Check Save Destination assignment assigned ...

Страница 116: ...ting assigned as desired by the customer Table 4 13 Peripheral Checks Step Task to do Expected Result s 1 Press the FREEZE key Stop image acquisition 2 Press the PRINT A or PRINT B key on the Control Panel The image displayed on the screen is printed on B W or Color printer depending on the key assignment configuration 3 Press the VCR key on the Control Panel twice VCR starts recording REC will be...

Страница 117: ...rd grasp it at the front grip of the user interface and rotate the console It is possible to rotate the Control Console up to 30 to the right WARNING WARNING Do not put your hand between the control console and the main unit when moving the console to the 0 position Danger of injuries Figure 4 30 Front Wheel with Brake Table 4 15 Brakes and Direction Lock Step Task Expected Result s 1 Flap the foo...

Страница 118: ... 4 7 Mechanical Function Checks 4 7 3 Power Supply Adjustment There are no adjustments on the power supplies The DC Power is self regulated If a voltage is outside the specified range it means that something is wrong either with the power supply itself or with a component connected to that specific power outlet ...

Страница 119: ...omatically 4 Enter the password SHE and click the ACCEPT button to display the Service Tools window 5 Click the MAINTENANCE REPORT button The following message box will be displayed 6 Fill in the requested information and click OK 7 Click the EXIT button on the Service Tools window and the EXIT button on the System Setup Service page NOTE After Hardware or Software modifications normally the Maint...

Страница 120: ...CAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL 4 36 Section 4 8 Site Log 4 8 2 Site Log Paper Documentation Table 4 16 Site Log Date Service person Problem Comments ...

Страница 121: ... and subsystem function It also describes the Power Distribution System PDS and probes Table 5 1 Contents in Chapter 5 Section Description Page Number 5 1 Overview 5 1 5 2 General Information 5 2 5 3 Main board Chassis GEF Module 5 18 5 4 Front End Processor 5 19 5 5 Back End Processor 5 24 5 6 Internal I O 5 28 5 7 Top Console 5 31 5 8 Monitor 5 32 5 9 External I O 5 33 5 10 Peripherals 5 34 5 11...

Страница 122: ... Scale and 2D Color Doppler Imaging CFM PD TD M Mode Gray Scale Imaging Doppler PW CW 3D Mode and Real Time 4D Imaging Different combinations of the above modes Figure 5 1 Voluson 730 Major Components Monitor Touch Panel Display TGC Slider Controls Monitor Controls Control Console Front Grip VCR Bolt to rotate and lock the Control Console mounted at the front below the Console Probe Connectors Foo...

Страница 123: ...ole User interface System I O with hard keys Touch screen and EL Display Section 5 7 on page 5 31 MONITOR Section 5 8 on page 5 32 GES External I O Connection Module Section 5 9 on page 5 33 GEM Removable Disk drive module ECG preamplifier MAN6 optional Section 5 10 on page 5 34 CPN Primary Power supply and Isolation transformer for the peripherals Section 5 11 on page 5 35 GW System mechanical ch...

Страница 124: ...lboard CPF Doppler CFM board CPM PC Motherboard Subboards Slot CPU several PC Cards CKV Video Card CPS DMA Controller Card CPE Backplane Card CPP Powersupply Secondaryboard GEU User Interface Module Monitor GES External IO GEF Main board Chassis CPN MainPowerSupply Module Subboards 32xCPD CPZ Coverboard CPK Motherboard of GEF Module GEM ECG Diskdrive Module GW Trolley housing mechanical framework ...

Страница 125: ...include Obstetrics Gynecology and Fertility Radiology Internal Medicine Neurology Cardiology Oncology Urology Orthopedics Pediatrics The system is designed for follow up expansion In addition to the initial operational settings for each transducer pre programmed in the system user customized parameter settings for each transducer may be inserted by the operator and stored for recall as needed via ...

Страница 126: ...y optional see 5 2 5 8 5 2 1 3 1 Color Flow Mode a real time two dimensional cross section image of blood flow is displayed The 2D cross section may be presented as a rectangle parallelogram trapezoid sector or a full circle depending on the particular transducer used The 2D cross section is presented as a full color display with various colors being used to represent blood flow velocity variance ...

Страница 127: ...30 Ultrasound System will be used to acquire multiple sequential 2D images which can be combined to reconstruct a three dimensional image These 3D images are useful in visualizing three dimensional structures and in understanding the spatial or temporal relationships between the images in the 2D sequence The 3D image is presented using standard visualization techniques such as surface or volume re...

Страница 128: ...on of a three dimensional object One is to rotate the volume of data and present the resulting sequence of 2D projections to the observer The changing direction of observation helps the observer to separate the features in the volume according to their distance from the observer 5 2 2 3 3D Rendering The 3D volume rendering is a calculation process to visualize certain 3D structures of a scanned vo...

Страница 129: ...16 CPY Powerswitch Plugboard CCF Powersupply Primaryboard CPH Trafo Console FootSwitch StdbySwitch User Power ON OFF CPN GEM ECG Diskdrive Module External Diskdrives ECG Module Aux_Supply SBC Single Board Computer Components Network USB Sound in old Version separate PCB SCSI in old Version separate PCB MajorTasks SystemControl 3D Rendering ImageProcessing CRI StrainImage etc Mouse RS232 SCSI 1 KVN...

Страница 130: ...ing 60MHz by 2 3 4 5 and TX Focus RX Channel ASIC generates Sample Clocks for the ADC manages RX Focus Delay and Chain Adder and Apodization 3 CPG Contains Multibeam DeInterleave Subtraction Filter for THI Digital TGC DC Canceler Mixer with NCO Numeric Controlled Oscillator FIR LP Low Pass Filter Decimation Magnitude Calculator Log Amp Re Sample Edge Enhance Line Filter Blending and Frame Filter T...

Страница 131: ...functions of CPG Image is composed of single at close range placed images PC calculated Only possible with 4D Probes 4 CPF Has no function in B Mode 5 CPS B mode Data from CPG are written into Line Memory Cine Mode Mem The DSC Dig ScanConverter reads out the Lines and writes them scan converted A mode polar to B mode cartesian into the Frame Memory From there the non deflected Image can be read ou...

Страница 132: ... Output to CPF 4 CPF Arranges the FFT D Mode Data use the same Bus to the CPS as CFM Data 5 CPS D Mode Data from CPF are written see 5 2 4 1 B Mode 6 CKV see 5 2 4 1 B Mode 5 2 4 4 CFM Mode Color Flow Mode 1 CPC see 5 2 4 1 B Mode 2 CPR see 5 2 4 1 B Mode 3 CPG After DC cancelling the sig is mixed with the NCO generated RX Frequency and brought to LF Spectrum where the FIR LP cuts HF Output to CPF...

Страница 133: ...emory Frame Memory Path is active in context with 3D Mode and Real Time 4D Mode just for displaying the actual scanned 3D Acquisition Sweep B Frame All other displayed B Frames in 3D 4D or SonoView Mode are PC generated and do not need CPS any more 6 CKV see 5 2 4 1 B Mode 5 2 4 6 Real Time 4D Mode 1 CPC see 5 2 4 1 B Mode 2 CPR see 5 2 4 1 B Mode 3 CPG see 5 2 4 1 B Mode 4 CPF Arranges the CFM Pr...

Страница 134: ...gnal Processing 5 CPS V Rec Data from the CKV Video Decoder are input to the CPS via the CFM Data Bus CPF must disable Output Driver In the LM FPGA the data are synchronized and output to the FM FPGA In the Extern Video Mode the FM FPGA is specially configured that s why the right part of the CPS Blockdiagram is not valid in this Mode In this Mode the FM FPGA re scales the Videorecorder Image that...

Страница 135: ... like biopsies ultrasound is a widely used method to visualize and guide the needle during puncture The advantage in comparison with other imaging methods is the real time display quick availability and easy access to any desired region of the patient The 4D biopsy allows for real time control of the biopsy needle in 3D multi planar display during the puncture The user is able to see the region of...

Страница 136: ...red on the Voluson 730 In addition Sonoview allows users to send and print images via DICOM network also Sonoview enables storage and retrieval of 2D Images and Cine Loops as well as 3D RT4D volumes and Live3D and RT4D sequences on HD Drive MO Drive and CD Drive 5 2 5 7 CRI Compound Resolution Imaging In this special B mode beams are transmitted not only perpendicularly to the acoustic window but ...

Страница 137: ...e steered allowing interrogation along an operator selected line within the image This option can be upgraded by implementing the CW Dopplerboard CRW old version CWD 5 2 6 2 ECG Digital Module The ECG preamplifier is used for acquiring an ECG signal to be displayed with the ultrasound image This optional peripheral serves for gaining an ECG signal to mark the systolic and end diastolic moments in ...

Страница 138: ...odule The GEF Module contains the Front End processor and the Back End processor and the Secondary Power supply for the full GEF Chassis Additionally GEF Module is the connection point of the internal I O wiring Figure 5 7 Mechanic of GEF Module Front End Secondary Power Board CPP Back End Front Rear with Internal I O Audio Video View from Right with Internal I O PC part Top View ...

Страница 139: ...isation RxFocusDelay Summation 8x TX PowerAmplifier 4x TGC AntiAliasingFilter ADC Each CPD capable of 8 TX Channels 4 RX Channels 32 x CPD CW Channels selected by Relais T R 256 CWTX 16 ExternVideo Data EBF_D 18 CW Data serial D Q 16Bit each CPU Control MCPU_A 12 MCPU_D 8 CPU Control MCPU_A 12 MCPU_D 8 ED_IN 16 EQ_IN 16 SIN COS SIN COS Powered CPG MID Processerboard For B C D modes Multibeam DeInt...

Страница 140: ...als CTRL signals CPU Probe Board Power 192 TX RX MOT_signals CTRL signals Power 192 TX RX MOT_signals CTRL signals TX RX Probe 1 Probe 2 Probe 3 CW Probe CPK Motherboard PowerSupply SIN COS Mot Strobes CTRL signals MCPU_A 0 12 MCPU_D 0 7 CPZ Cover Board 32 CPD Beamformer Subboard CPR Beamformer Motherboard Transmitter Receiver to CW_Board CWD newlyCRW 192 Tx Rx Channels 32 Tx Rx Channels EBF_D 18 ...

Страница 141: ...es TX Pulses TX Focus TX Apodisation Rx Focus Delay and Summation 8x TX Power Amplifier 4x TGC Anti Aliasing Filter Analogue Digital Conversion ADC Each CPD is capable of 8 TX Channels and 4 RX Channels 5 4 1 3 CRW old version CWD CW Doppler Board Optional Separate Clock Oscillator Clock generator PRF FN TxDelayMatrix Transmitter Receiver Amplifiers RxAmp Receiver Delay matrix RxDelayMatrix Analog...

Страница 142: ...ending Frame Filter 5 4 1 6 CPC Control Board Clock Oscillator 60MHz PRF Pulse Repetition Frequency Generator Beamformer Configuration Tx transmitting Frequency TxRx transmitting receiving Focus timing TxRx transmitting receiving Apodisation Line number lateral Position contains TX Power DAC 5 4 1 7 CPF Doppler CFM Board A CFM processing Sample rate Conversion Auto Correlator Wall motion filter WM...

Страница 143: ...ply of both Front End and Back End DC DC Converter 59VDC to following output voltages 3 3V 5V 12V 15V FAN 24V DC TX_POW 90V fused by F7 TX_POW 90V fused by F1 Motor Sinus2 Powerstage Amplifier fused by F3 Monitor Cosinus2 Powerstage Amplifier fused by F6 5 4 1 10 CPK Motherboard of GEF Module Following boards are direct connected to the CPK CPU BYM Probe connector Board CPR Beam former Board CRW C...

Страница 144: ...rlay Logic VGA Adapter Video Converter RGB VHS etc CPY Powerswitch Plugboard CCF Powersupply Primaryboard CPH Trafo Console FootSwitch StdbySwitch User Power ON OFF CPN GEM ECG Diskdrive Module External Diskdrives ECG Module Aux_Supply SBC Single Board Computer Components Network USB Sound in old Version separate PCB SCSI in old Version separate PCB MajorTasks SystemControl 3D Rendering ImageProce...

Страница 145: ...a RAM I6 DMA Descriptor RAM 128x32 Xilinx Blockram Local Bus I F Ext Bus I F DMA Controller PCI 9054 I5 EEPROM Clk Gen Quad UART ST16C554 I16 RS232 Driver Receiver I21 3 x 74ALVCH16601 LCX574 I4 I7 I8 I11 PCI Bus 33MHz 66MHz RS232 VCR Remote 33MHz CFM Slot BW Slot RD Slot SDRAM Contr BW CFM RD Slot Count Enables DSC_LA0 DSC_LA1 TVX TVY SDRAM Control CLKEN VGA HSYNC RED GRN BLU to CKV EAD Ctrl PC C...

Страница 146: ...onverter for 2D DMA Controller f High Speed Data Transfer into Slot CPU Memory 5 5 2 3 CKV Video Card Overlay Logic Adding the Overlay graphics to the Ultrasound graphics VGA Output 2 Channels for the System Main Monitor Video Converter RGB VHS External Video Playback PAL NTSC Video Generation etc 5 5 2 4 Hard Disk Drive Minimum 20GB IDE Stores the system programs and Image filing patient data Rep...

Страница 147: ... End Board Descriptions cont d 5 5 2 7 CPP Power Supply Secondary Board Motor Power stage used for Supply of both Front End and Back End DC DC Converter 59VDC to following output voltages 3 3V 5V 12V 15V FAN 24V DC TX_POW 90V fused by F7 TX_POW 90V fused by F1 Motor Sinus2 Powerstage Amplifier fused by F3 Monitor Cosinus2 Powerstage Amplifier fused by F6 ...

Страница 148: ...50MHZ PC separate Sound card separate SCSI Card installed Figure 5 13 Internal I O Board 2 3 SNO s A06269 A 1GHZ PC Sound onboard SCSI onboard Figure 5 14 Internal I O Board 3 Figure 5 12 Internal I O Board 1 CPN FootSwitch External IO Board IN OUT VGA R G B Sync NET USB L AUDIO R VHS S VHS VGA 1 VGA 2 RGB S VHS 1 S VHS 2 VHS 1 VHS 2 VHS S VHS AUDIO OUT IN SOUND IN R L CKV CPE GEM 50 pol Monitor P...

Страница 149: ...N FootSwitch External IO Board IN OUT VGA R G B Sync NET USB L AUDIO R VHS S VHS VGA 1 VGA 2 RGB S VHS 1 S VHS 2 VHS 1 VHS 2 VHS S VHS AUDIO OUT IN SOUND IN R L CKV CPE GEM 50 pol Monitor PC USB PS2 RS232 NET 50 pol KVR ECG RS232 CONSOLE REMOTE BW PRINTER KVG KVA AUX SUPPLY LAMP CPN V SYNC not conn SCSI StdbySwitch User PowerON OFF LINE OUT MIC IN LINE IN GEU User Interface Modul ...

Страница 150: ...N OUT VGA R G B Sync NET USB L AUDIO R VHS S VHS VGA 1 VGA 2 RGB S VHS 1 S VHS 2 VHS 1 VHS 2 VHS S VHS AUDIO OUT IN SOUND IN R L CKV CPE GEM 50 pol Monitor USB PS2 RS232 50 pol KVR ECG RS232 CONSOLE REMOTE BW PRINTER KVG KVA AUX SUPPLY LAMP CPN V SYNC not conn USB LAN1 LAN2 LINE OUT RS232 Line In MIC USB PC PS2 CPN FootSwitch StdbySwitch User PowerON OFF GEU User Interface Modul ...

Страница 151: ...e module Embedded 80C32 console micro controller Built in serial UART for PC104 communication Slide pots TGC with zero raster position Rotary Encoders with integrated push buttons PS 2 compatible Trackball 2 with standard PC interface PS 2 compatible Qwerty Keyboard with standard PC interface LED indicators with 3 intensity levels off 50 100 2 Speaker used for Doppler and voice replay DC DC Conver...

Страница 152: ...CHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL 5 32 Section 5 8 Monitor Section 5 8 Monitor For further details refer to Section 6 3 Monitor Adjustment on page 6 2 Figure 5 16 Monitor Adjustment buttons ...

Страница 153: ...I O may not be available Please refer to Section 5 6 Internal I O on page 5 28 Table 5 4 External I O Connector Description Item Connector Name Description 1 VGA OUTPUT print out VGA signal with monitor printer 2 R G B H V SYNC outputs for color video printer monitor 3 NETWORK DICOM input output twisted pair RJ 45 10 100 megabit s 4 USB 1 USB port 5 USB 2 USB port 6 AUDIO LEFT OUT 7 AUDIO RIGHT OU...

Страница 154: ...nitor and must not be used for an intra operative application on the heart 5 10 1 2 CD RW Drive Internal SCSI 24x Read 8x Write 8x Re Write CD RW Drive Access Speed 140ms Max Data Transfer Read 3600KB sec Buffer size 4MB Recordable and Re writable CD s are ideal for any storage purpose and offer complete security and reliability for important data 5 10 1 3 Magneto Optical Drive Storage capacity by...

Страница 155: ...rear of the system The Mains Switch is of the auto fuse type if for some reason the current grows to high the switch will automatically break the power From the Mains Switch the AC power is routed via an Inrush Current Limiter to a internal outlet connector for the Mains Transformer 5 11 1 2 Major Functions of CPN Inrush Current limiter Power factor correction transformer for Sinus load for the ma...

Страница 156: ...ker 2 5 5A F3 F4 F5 F6 are the fuses for the input voltage for the switching power supplies generating the DC Supply voltage for the Secondary power supply inside GEF Module 5 11 1 4 Fuses inside CPN F1 on CCF board fuses the surge current limiter circuit NOTE If this fuse is blown the NTC limiting the surge current will remain hot during system operation and if the system is switched off on withi...

Страница 157: ...RECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL Chapter 5 Components and Functions Theory 5 37 Section 5 12 Mechanical Descriptions 5 12 1 Physical Dimensions Figure 5 20 Physical Dimensions 1445 94 1067 50 851 75 667 50 607 61 677 77 561 25 ...

Страница 158: ...tor Tilt 10 forwards and backwards Swivel 45 rotation 5 12 3 Top Console Positioning 5 12 4 Rotation of the Control Console Horizontal Access The control panel offers 30 of horizontal adjustment to the right Figure 5 21 Top Console Positioning Figure 5 22 Locking Bolt under Control Console 30 Locking Bolt for locking and unlocking the control console ...

Страница 159: ...STEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL Chapter 5 Components and Functions Theory 5 39 5 12 5 Assembly Drawing GW GEU Monitor Figure 5 23 Assembly Drawing GW GEU Monitor ...

Страница 160: ...to the scanner through the filter grid and pushes it through the card rack Air holes in the left and right side of the rack allow the air to move down through the card rack Air exits the scanner through holes the Main Air Outlet at the left side panel 5 13 1 1 Air Flow Distribution Overview Figure 5 24 Console Views Figure 5 25 Air Flow Control Main Air Outlet CPN Air Outlet Main Air Inlet Filter ...

Страница 161: ...ce productivity and reduce training and service costs 5 14 2 Access Security The service page has different access and security user levels Each user is only granted access to the tools that are authorized for their use 5 14 3 Service Login 1 On the Touch Panel press UTILITIES 2 In the Utilities menu touch SYSTEM SETUP to invoke the setup desktop on the screen 3 Select the SERVICE page The passwor...

Страница 162: ...ns scanning Touch Panel entries performing Calculations review of Patient Reports etc It is possible to safe record as file on HDD But also export to CD or MOD can be done to allow replay of the records on other units Operation see Section 7 5 How to use the Auto Tester program on page 7 5 Figure 5 27 Service Tools window NOTICE For intermittent problems this file can be requested from the Service...

Страница 163: ...ware Options and Service Actions to MO or CD Drive 5 14 3 4 Export Event Log Select this button to export the Event Log File to MO or CD Drive 5 14 3 5 Delete all Patients 1 Click the DELETE ALL PATIENTS button Following WARNING message appears on the screen 5 14 3 6 Maintenance Report Any modification upgrade and maintenance action should be entered in this report to get a history of all service ...

Страница 164: ...matically on the screen 5 14 3 7 Export System Data Select this button to transfer the Service Database to MO or CD Drive 5 14 3 8 Update 5 14 3 8 1 EUM is for updating the Electronic User Manual Operation see Section 8 3 EUM Electronic User Manual Upgrade Procedure on page 8 3 5 14 3 8 2 NLS Not for use in the field 5 14 3 9 Printer Installation of Printers is possible without entering the Window...

Страница 165: ... specification Although some tests may be optional they should only be performed by qualified personnel Section 6 2 Regulatory Verify where applicable that any regulatory information or tests required by national law are present and accounted for and any regulatory tests required by national law are performed and documented Table 6 1 Chapter 6 Contents Section Description Page Number 6 1 Overview ...

Страница 166: ...n 6 3 Monitor Adjustment Color temperature Geometry Size Picture location Brightness and Contrast are adjusted using the keys located on the front of the monitor just below the screen The screen is degaussed automatically when the power is turned on Figure 6 1 Monitor Adjustment buttons Figure 6 2 Description of Monitor Adjustment buttons ...

Страница 167: ...ect UTILITIES SYSTEM SETUP and click the DATE TIME button on the General page and look at background of calender GEOM no pincushion distortion no picture rotation Monitor is pre adjusted minor distortion is normal and can be corrected with the GEOM function SIZE video display fills up full screen CENT video centered left and right margin equal Table 6 3 Brightness Contrast Settings Room Condition ...

Страница 168: ...sole to wanted position 6 4 2 Transport Lock There is a locking bolt for locking and unlocking the control console mounted at the front below the control console When preparing the system for transport the lock has to be engaged in order to secure the console against uncontrolled rotation The lock catches in when the console is rotated to its center 0 position Figure 6 3 Locking Bolt under Control...

Страница 169: ...ng by turning it counterclockwise 2 Adjust the trackball for smooth running by rotating the black securing ring CCW smooth run CW tighten run 3 Mount the outer fixing ring by turning it clockwise see Figure 6 4 Figure 6 4 Trackball with fixing ring Figure 6 5 Trackball Securing Ring NOTICE Avoid tightening of thread caused by improper mounting of securing ring rotate counterclockwise to remove the...

Страница 170: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL 6 6 Section 6 5 Trackball Adjustment This page was intentionally left blank ...

Страница 171: ...the application level 7 1 2 Overview There may be a time when it would be advantageous to capture trouble images and system data logs for acquisition through remote diagnostics InSite or to be sent back to the manufacturer for analysis There are different options to acquire this data that would give different results Table 7 1 Contents in Chapter 7 Section Description Page Number 7 1 Overview 7 1 ...

Страница 172: ...alyze data or images being reported as a malfunction or being returned to the manufacturer 1 System Serial number External number Label on back of the system Internal number Touch UTILITIES then SYSTEM SETUP and select the SYSTEM INFO page 2 Applications Software Touch UTILITIES then SYSTEM SETUP and select the SYSTEM INFO page NOTICE Voluson 730 can have two different serial numbers Figure 7 1 Sy...

Страница 173: ...eck power Figure 7 2 Connectors Pin 15 5V Pin 13 15V Pin 11 15V ECG Check above Voltage with DVM to Ground DVM Check with DVM Digital Volt Meter the 58V DC Check above Voltages with DVM to Ground 1 8 9 15 Backpanel of GEF SCSI Power Connector Backpanel of GEF Power Connection Cable from CPN to GEF 12V Pin 1 1 2 3 Check Voltage with DVM to Ground both Pins 2 and 3 are 3 3V Pin 1 is Ground ...

Страница 174: ...c process to capture any screen from the scanner 1 Navigate to and display the image screen to be captures 2 Press SAVE button and save the image to SonoView 3 Select the stored image s in SonoView Exam List and export the image s to CD Rom jpg bmp or tiff 7 4 2 Export Log s and System Data s Touch UTILITIES then SYSTEM SETUP and select the SERVICE page Type in the password SHE and click the ACCEP...

Страница 175: ...ting 7 5 Section 7 5 How to use the Auto Tester program 1 Touch UTILITIES and SYSTEM SETUP on the Touchscreen 2 Select the SERVICE page on the screen 3 Type in the password SHE and click ACCEPT The following screen appears 4 Activate the Auto Tester program by clicking START Figure 7 4 System Setup Service Figure 7 5 Service Tools ...

Страница 176: ... the Record icon on the displayed screen 8 Start scanning You can scan normally and everything will be recorded to the program up to several hours NOTE It is important that you are recording the processes where the errors normally occur 9 Stop the program by pressing the PAUSE BREAK key on the right upper corner of the keyboard Figure 7 6 Message Box Figure 7 7 Start Auto Tester Pause Break key on...

Страница 177: ... screen will appear 10 Insert an empty CD RW in the Drive and select the CD Burn icon 11 Enter a Filename 12 After clicking OK the following message boxes will appear 13 After the CD write is finished click the OK button and close the Auto Tester program Figure 7 8 Auto Tester Finished Figure 7 9 Enter a Filename Figure 7 10 CD Burn Process CD Burn icon ...

Страница 178: ...bles must be connected for minimum Configuration see Figure 7 11 Cable Minimum configuration a RS232 1 connector for User Interface near to the Mouse connector NOTE Don t mix up the RS232 1 with the RS232 2 and RS232 3 connectors b VGA Monitor c CPN Primary Power Supply d Console e Stand By switch f Monitor Power Supply g Mouse connector for Mouse Keyboard Figure 7 11 Cable Minimum configuration g...

Страница 179: ...n Following Cables must be connected for minimum Configuration see Figure 7 11 Cable Minimum configuration a RS232 1 connector for User Interface near to the Mouse connector NOTE Don t mix up the RS232 1 with the RS232 2 and RS232 3 connectors b VGA Monitor c CPN Primary Power Supply d Console e Stand By switch f Monitor Power Supply g Mouse connector for Mouse Keyboard h Probes ...

Страница 180: ...st items checked if the scanner doesn t boot up If all three conditions are present the scanner should be in astandby mode with the ON OFF switch illuminated with amber color Follow the flow chart to determine the reason the scanner doesn t boot up No No Switch in ON Position Fan Noise Yes Check if the power switch over the mains plug of the Primary Power Supply isn t switched off Check the connec...

Страница 181: ...0 SERVICE MANUAL Chapter 7 Diagnostics Troubleshooting 7 11 7 7 2 Noise in Image Figure 7 13 Noise in Image Noise is shown in Image YES NO Is there any electrical devices that affect the Scanner Keep the Scanner away from those electrical devices use another wallsocket Check probes ...

Страница 182: ...SON 730 SERVICE MANUAL 7 12 Section 7 7 Troubleshooting Trees and Instructions 7 7 3 Trackball Figure 7 14 Trackball Troubleshooting The trackball is low sensitive Remove the ball from the Trackball and clean it YES NO Does the trackball move at all Replace the trackball unit ...

Страница 183: ... 7 4 System does not Power Off Shutdown Figure 7 15 Power Off Shutdown Troubleshooting System has been switched off System starts with the shut down procedure System is down Yes system switched off automatically in app 3min Primary Power CPY inside Supply is defekt check the switch and the cable if O K Primary Power Supply is defekt No Yes No ...

Страница 184: ...rm Turn on the power to activate the Auto Degauss function Screen image is not centered or sized properly Adjust the picture location picture size picture rotation or pincushion distortion Some video modes do not fill the screen to the edge of the monitor There is no single answer to solve the problem This phenomenon may occur on higher refresh rates vertical frequency Picture is fuzzy Adjust the ...

Страница 185: ...Unable to record to VCR Check that all signal and power supply cable connections to the VCR unit are OK YES NO Is the VCR properly connected Is the VCR tape put into the device Put the tape into the device and rewind it NO YES Replace VCR Is recording impossible even by manual operation YES Check the signal cable connections between VCR and Internal I O NO YES Video Output of System o k Replace CK...

Страница 186: ...ter Troubleshooting Figure 7 18 Printer Troubleshooting Unable to use Printer s Check that all signal and power supply cable connections between Printer and Scanner are OK YES NO Is the printer s properly connected Are there print papers left Insert the printer papers NO YES Is proper configuration for the Print key s set Configure the Print key s NO See next page ...

Страница 187: ...17 7 7 7 Printer Troubleshooting cont d Figure 7 19 Printer Troubleshooting cont d Still unable to use the Printer s Check the signal cables between GEF and Backpanel YES NO Is printing unavailable even by manual operation Still no image on the print out Video Output of System o k NO YES Printer o k Replace the Printer Replace CKV NO YES ...

Страница 188: ...ng backup should be visible 7 7 8 MOD Troubleshooting 1 Insert an empty MO into the Drive 2 Enter Sonoview by pressing the NETWORK key on the control panel see Figure 7 20 3 Click the Open icon to display the list of exams 4 Select exam s and backup them to MO 5 Choose MO Drive 6 The images which you have chosen during backup should be visible 7 7 9 Audio Test Preparations for Audio Test Start a p...

Страница 189: ...eck that the network cable between the scanner and the wall network is connected and well seated in both ends 2 Try a network cable that is known to be OK 3 Check the cable between the network connector on the Back Panel to the LAN connector on the GEF 4 Connect a network cable between your Scanner and your PC Try to ping from the Scanner to the IP address on the PC If OK the hardware connection i...

Страница 190: ...Of Range Press ok and save this user setting once again BC_Lines BC_LineDensity out of limit restart the system BCMC_Balance Out Of Range Press ok and save this user setting once again BC TxPower Out Of Range Press ok and save this user setting once again BF can t set BM RxApod restart the system If after restart not ok then call technical support BF can t set C Rx Apod restart the system If after...

Страница 191: ...ct the probe once more if not ok call technical support Cannot write Start Date into EEPROM on left probe connector B connect the probe once more if not ok call technical support Cannot write Start Date into EEPROM on left probe connector C connect the probe once more if not ok call technical support CANNOT_IMPORT_VOLUME_DATA_TO_3D_DLL Load volume files from other storage medium Can t import sessi...

Страница 192: ...g once again End Bandwidth too big restart the system If after restart not ok then call technical support End ET too big restart the system If after restart not ok then call technical support End frequency too big restart the system If after restart not ok then call technical support Enhance Out Of Range Press ok and save this user setting once again Ensemble Out Of Range Press ok and save this us...

Страница 193: ...e medium File Not Found Load another Backup from CD restart or reinstall the Application File Not Found restart install latest software version File Pos Load volume files from other storage medium File Read Error Load volume files from other storage medium File Type Unknown Load volume files from other storage medium File Volume size not consistent Load volume files from other storage medium FLT D...

Страница 194: ...rm the message by pressing the OK button disconnect and connect the active probe and use it again if again the message appears mostly the probe itself has an defect so you should contact the service MotCtrlDrv bCmdToRS232 restart the system M TxPower Out Of Range Press ok and save this user setting once again No CD Writer found check the SCSI and the Power cable plug the SCSI cable off and on and ...

Страница 195: ...t Of Range Press ok and save this user setting once again Start Bandwidth too small restart the system If after restart not ok then call technical support Start ET too small restart the system If after restart not ok then call technical support Start frequency too small restart the system If after restart not ok then call technical support System detected severe error please call technical support...

Страница 196: ...own CPF_1 Xilinx Version restart the system unknown CPF_2 Xilinx Version restart the system unknown CPG1 Xilinx Version restart the system unknown CPG2 Xilinx Version restart the system unknown CPG3 Xilinx Version restart the system Unknown Error Load volume files from other storage medium unknown FMV Xilinx Version restart the system unknown LMV Xilinx Version restart the system unknown Xilinx Ve...

Страница 197: ...or example body coils or and ultrasound probe The purpose of the regulation is to protect employees in the transportation industry as well as the people who will receive or open this package The US Department of Transportation DOT has ruled that items what were saturated and or dripping with human blood that are now caked with dried blood or which were used or intended for use in patient care are ...

Страница 198: ...ding Procedure Section 8 2 Backup Loading Procedure The Backup contains User settings Auto Text Measurement user tables Biopsy needle guide positions Setup settings language date format screensaver on off etc 8 2 1 Loading Procedure see Section 4 4 12 3 Load System Presets and Configurations Application Settings on page 4 28 ...

Страница 199: ...7 Click the EUM button The Update Software dialog appears 8 Select CD DRIVE Browse for the EUMSetup_en exe file in the Install folder and click OK 9 Confirm the following message with YES to proceed the update 10 The contents of this package are being extracted 11 Follow the instructions of the InstallShield Wizard confirm the messages with the NEXT key NOTICE The Loading Procedure of the EUM Elec...

Страница 200: ... electronic user manual does not work till the next system start 12 Remove any disks from their drives and click FINISH to complete the EUM installation The system will reboot automatically NOTE If the End Program message box appears select END NOW 13 After rebooting the system fill in the requested information into the System Status Messenger box and click OK 14 Press the F1 key on the keyboard t...

Страница 201: ...ke the setup desktop on the screen 4 Select the SERVICE page enter the password SHE and click the ACCEPT button The Service Tools menu appears on the screen 5 Click the EUM button The Update Software dialog appears 6 Select the MO DRIVE or the CD DRIVE button depending on the storage medium you use 7 Browse for the EUMSetup_en_xxxx exe file and click OK NOTICE The electronic user manual EUM must n...

Страница 202: ...m the next message with YES to proceed the update The contents of the package are being extracted 9 Read the following warning message and confirm it with OK 10 Confirm the Maintenance Complete message InstallShield Wizard has finished performing maintenance operations on V730 User Manual with the FINISH button Figure 8 5 confirm warning message NOTICE At this step the old EUM files will be uninst...

Страница 203: ...this package are being extracted 5 Follow the instructions of the InstallShield Wizard confirm the messages with NEXT 6 After successfully installation of the Electronic User Manual click FINISH to exit the wizard 7 Close the SERVICE page with EXIT and close the System Setup with SAVE EXIT 8 Restart the system turn off and on the system 9 After rebooting the system fill in the requested informatio...

Страница 204: ...placement of the Trackball top fixation ring 8 4 1 Manpower One person 5 min 8 4 2 Trackball top fixation ring Replacement Procedure 1 Remove the fixation ring by turning it counterclockwise 2 Mount the fixation ring by turning it clockwise Figure 8 7 Trackball with top fixation ring rotate counterclockwise to remove the fixing ring rotate clockwise to mount the fixing ring ...

Страница 205: ...le for the Voluson 730 Table 9 1 Contents in Chapter 9 Section Description Page Number 9 1 Overview 9 1 9 2 List of Abbreviations 9 2 9 3 Parts List Groups 9 3 9 4 Housing GW and additional Console Hardware 9 4 9 5 User Interface GEU Monitor 9 6 9 6 Disk Drives GEM 9 9 9 7 Main Power Module CPN 9 10 9 8 Main Board Module GEF 9 11 9 9 Options and Upgrades 9 18 9 10 Miscellaneous Cables 9 19 9 11 Op...

Страница 206: ...in parts hub FRU 2 Replacement part available from the manufacturer lead time involved GW Trolley including Console and Monitor mounting plate except GEU GEU User interface Keyboard Touch Screen EL Display TGC Unit GEM Disk Drive module with or without MAN CPN Primary power module GEF Main Board Module Ultrasound FrontEnd and PC Boards Backend Processor MAN ECG module GES I O Interface User access...

Страница 207: ...er module 500 510 520 530 570 Main Board Module GEF FrontEnd US Part FrontEnd US Part cont d FrontEnd US Part cont d BackEnd Processor PC Part Table 9 8 on page 11 Table 9 9 on page 12 Table 9 10 on page 14 Table 9 11 on page 15 Table 9 12 on page 17 GEF Main Board Module Ultrasound FrontEnd and PC Boards BackEnd Processor 600 Options and Upgrades Table 9 13 on page 18 700 Miscellaneous Cables Tab...

Страница 208: ...ERVICE MANUAL 9 4 Section 9 4 Housing GW and additional Console Hardware Section 9 4 Housing GW and additional Console Hardware Figure 9 2 Housing GW and additional Console Hardware 101 102 103 104 105 106 107 108 109 110 111 112 113 114 115 116 117 118 119 120 121 122 123 124 125 126 ...

Страница 209: ...730 1 1 111 Monitor Mounting Plate from A05722 KTZ207270 Monitor mounting plate incl probe holder 1 1 112 Monitor Mounting Plate up to A05721 KTZ154604 Monitor mounting plate 1 1 113 Probe Holder up to A05721 KTZ13K815 Probe holder for Vaginal probe 1 1 114 Rear Wheel Ø175 mm x 32 KTZ211081 Rear wheel non steerable 1 1 115 Blind Cap for Wheel KTZ14M871 Cover for the rear wheels 1 1 116 Steerable W...

Страница 210: ...eserved for V530D Please order KTZ211096 1 1 202 Color Monitor 15 Painted KTZ211096 Sony PGM 100P1MD KR lacquered 1 1 203 Color Monitor 15 Painted KTZ211212 Sony PGM 100P1MD KR lacquered slightly different color shade can replace KTZ211096 1 1 204 Monitor mount Shell KTZ208509 fixing part for Sony Monitor 1 1 205 Front Handle User Interface KTZ220039 V730 Front Handle 1 1 206 Probe Holder Kit V730...

Страница 211: ...SOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL Chapter 9 Renewal Parts 9 7 Section 9 5 User Interface GEU Monitor cont d Figure 9 4 User Interface GEU Monitor cont d 207 208 209 210 211 212 214 215 216 217 218 219 220 221 213 ...

Страница 212: ...2 GEU4 6 PC Keyboard kit KTZ207115 GEU4 6 PC alphanumeric Keyboard kit 1 1 213 Left PWB kit for V730 KTZ207108 Printed Wire Board inside GEU 1 1 214 Right PWB kit for V730 KTZ207109 Printed Wire Board inside GEU 1 1 215 Slide Potentiometer kit for V730 KTZ207110 TGC Slider controls board inside GEU 1 1 216 GEU1 2 User Interface AID without PS II KTZ154621 keyboard trackball display special knobs s...

Страница 213: ...g of GEM SCSI Drive Module 1 1 303 GEM1 1a SCSI Disk Drive Module KTZ195474 Mechanical Module for SCSI Drives MO and CD RW used in serial production up to A06310 1 2 304 GEM2 MO Drive CR RW KTZ154652 Mechanical Module for SCSI Drives MO and CD RW replaces GEM1 1a used in serial production from A06311 1 2 305 GEM3 3a MO Drive CR RW KTZ154673 Mechanical Module for SCSI Drives MO and CD RW replaces G...

Страница 214: ...02 CPN4 Power Supply Module KTZ195564 CPN4 4 Power Supply Module replaces CPN3 used in serial production from A05801 A06353 1 1 403 CPN5 Power Supply Module KTZ195739 CPN5 Power Supply Module replaces CPN4 used in serial production from A06354 1 1 404 CPY3 P3 Power Switch KTZ195471 Board to switch on and shut down Ultrasound System 1 1 405 Mains Power Input connector KTZ207574 Mains Power Input co...

Страница 215: ...Name Part Number Description Qty FRU 501 GEF1 2d Main Board Chassis KTZ195400 GEF1 2d Chassis mech electr Main Board used in serial production up to A06353 1 N 502 GEF3 Main Board Chassis KTZ195700 GEF3 Chassis mech electr Main Board can replace GEF1 2d used in serial production from A06354 1 N 503 Fan for GEF box 2 fan KTZ154678 Fan for GEF box 2 fan 1 1 504 Fan for GEF box single KTZ154679 Fan f...

Страница 216: ...ormer Board from A06354 KTZ195645 Beamformer Board can not be replaced by CPR4 4b P4 1 1 514 CWD2 CW Doppler Board KTZ195663 CW Doppler Board 1 1 515 CRW2 P2 CW Doppler Board KTZ195723 CW Doppler Board can replace CWD2 minimum software requirement 1 06 1 1 516 CPZ2 2a P2 Cover board up to A06353 KTZ195406 electrical signal connection between CPU CPR and CW Board contains Relays for Signal to CW Bo...

Страница 217: ...KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL Chapter 9 Renewal Parts 9 13 9 8 1 FrontEnd US Part cont d Figure 9 9 FrontEnd US Part cont d 521 522 523 524 525 526 527 528 529 530 531 532 ...

Страница 218: ...em Control Board from A06105 KTZ195654 System Control Board can not be replaced by CPC2 2f P2 1 1 525 CPC5 P4 System Contorl Board from Ap6105 KTZ195821 System Control Board can replace CPC4 P4 can not replace CPC2 2f P2 1 1 526 CPF3 3a P3 Color Doppler Board up to A06104 KTZ195439 Color Doppler Board can not be replaced by CPF10 10a P4 and CPF11 P5 except 1 1 527 CPF10 10a P4 Color Doppler Board ...

Страница 219: ...lace CPP2 2e P2 1 1 533 CPP4 4b P3 Power Supply Board KTZ195679 Power Supply Board Output Power 900 W can replace CPP2 2e P2 and CPP3 3c P3 1 1 534 CPP5 P3 Power Supply Board KTZ195826 Power Supply Board Output Power 900 W can replace CPP2 2e P2 CPP3 3c P3 CPP4 4b P3 1 1 535 CPU1 1d P1 Probe Connector Board up to A06353 KTZ195417 Probe Connector Board Module Board CW connector not fitted used in s...

Страница 220: ...K ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL 9 16 Section 9 8 Main Board Module GEF 9 8 2 BackEnd Processor PC Part Figure 9 11 BackEnd Processor PC Part 571 572 574 575 576 577 578 579 580 581 582 583 584 573 ...

Страница 221: ... 4a P4 Motherboard Extension KTZ195542 Motherboard Extension used in serial production from A06041 can replace CPE3 P3 1 1 577 CPM2 P2 Motherboard up to A06104 KTZ195191 electrical Signal and Supply Connection between all PC Plug In Boards can not be replaced by CPM3 P2 except 1 1 578 CPM3 P2 Motherboard from A06105 KTZ195699 electrical Signal and Supply Connection between all PC Plug In Boards ca...

Страница 222: ...Upgrades Item Part Name Part Number Description Qty FRU 601 MO Disk Media 1 3GB KTZ207077 MO Disk Media Standard 1 1 602 Service Boot CD Rom PIII 1GHz KTZ1C2192 Repair CD Rom Linux based only for 1GHZ SBC boards which have no GE licensed BIOS 1 1 603 PC Replace Kit PIII 1GHz up to A06267 KTZ195718 PC Board replacement for 850MHz 1 1 604 PCW1 1a CW Upgrade Kit from A05900 KTZ195615 CW Doppler Upgra...

Страница 223: ... Table 9 14 Miscellaneous Cables Item Part Name Part Number Description Qty FRU 701 Adapter SCSI II DB50 DB50 KTZ207121 SCSI Link 1 1 702 Adaptor 26Pin SUB D25 KTZ207085 Adaptor LPT1 Parallel Port 1 1 703 Adapter Line in MIC GEU KTZ207390 Serial MIC in Speaker out 1 1 704 Cable 50Pin SCSI 2 Data Cable 1 8m KTZ212403 Data Transfer between PC and SCSI Module 1 1 ...

Страница 224: ... Data Cable HDD KTZ212398 Data cable for IDE Harddisk 40pin 330mm 1 1 708 GES2 I O Connection Panel KTZ195503 External Rear Panel with electrical Signal and Supply Connection Cables to the V730 Main Unit internal Rear Panel 1 1 709 GES3 I O Connection Panel KTZ195638 External Rear Panel with electrical Signal and Supply Connection Cables to the V730 Main Unit internal Rear Panel 1 1 710 KGS10 Cabl...

Страница 225: ...195416 electrical Signal and Supply Connection between V730 and User Interface GEU 1 1 714 KVR2 TERM RS232PC V730 Y Cable KTZ207353 electrical Signal and Supply Connection between Voluson 730 and User Interface GEU can replace KVR1 cable KTZ195416 1 1 715 KVS2 Cable Serial Interface KTZ195717 electrical Signal Connection between CPS and PIC_MG Slot CPU 1 1 716 KVX1 Network Cable KTZ212016 Cable fr...

Страница 226: ...ne cord for Supply from Main Device to Sub Devices 1 1 718 Monitor Signal Cable RGB 75E KTZ212010 Monitor cable RGB 75E 2m VGA SVGA 1 1 719 PS2 Y splitter KTZ212440 Splits the PS2 connector for GEU4 6 mouse and PC keyboard 1 1 720 Power Cable for Hard disk KTZ212401 Power cable for Hard disk 4pin 230mm 1 1 Table 9 14 Miscellaneous Cables Item Part Name Part Number Description Qty FRU ...

Страница 227: ...rope 230V KTZ212317 Power Cord Europe 230V 240V 1 1 722 Power Cord Japan Hosp grade KTZ212448 Power Cord Japan Hospital Grade 1 1 723 Power Cord UK KTZ212441 Power Cord United Kingdom 240V 1 1 724 Power Cord USA Hosp grade KTZ212402 Power Cord USA Hospital Grade 1 1 Table 9 14 Miscellaneous Cables Item Part Name Part Number Description Qty FRU ...

Страница 228: ...ide 68pol Connector on PC Backpanel MD 50P female 1 1 726 SCSI Adapter 68 Pol 50 Pol KTZ212103 Leads the SCSI Signal from the PC Board to the PC Backpanel Connector on PC Side 68pol Connector on PC Backpanel MD 50P female 1 1 727 USB to ICP CPU Board KTZ207029 USB for PC Slot Connector on Backpanel Cables are connected to PC Board Leads the USB signals to the PC Backpanel 1 1 Table 9 14 Miscellane...

Страница 229: ... DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL Chapter 9 Renewal Parts 9 25 Section 9 11 Optional Peripherals and Accessories Figure 9 13 Optional Peripherals and Accessories 801 802 803 804 805 806 807 808 809 810 811 812 813 814 815 ...

Страница 230: ...on Set KTZ195776 Connection Set UP D21MD 1 1 806 Sony Digital Color Printer UP D2600S KTZ211060 SCSI Port 1 1 807 Mitsubishi Digital Color Printer CP770DW KTZ211352 SCSI Port 1 1 808 PZP50 Connection Set KTZ195486 Connection Set UP D2600S CP770DW 1 1 809 Sony VCR SVO 9500 MD KTZ211317 NTSC 1 1 810 Sony VCR SVO 9500 MDP KTZ211318 PAL 1 1 811 PRR50 Connection Set KTZ195492 Connection Set for VCR wit...

Страница 231: ... Number Description Qty FRU 816 ECG preamplifier MAN 3 KTZ154258 consists of ECG preamplifier and patient connection cable 1 1 817 ECG preamplifier MAN 6 KTZ154644 consists of ECG preamplifier and patient connection cable 1 1 818 Foot switch MFT7 KTZ195446 Foot switch 1 1 819 USB RS232 Connection kit PRY KTZ195858 Converter from USB to RS 232 Serial Port 1 1 820 Touch up Paint Set gray blue KTZ154...

Страница 232: ...y FRU Voluson 730 Service Manual KTZ105844 105844 1 N System User Manuals Basic User Manual Voluson 730 English H46601B 1 N Instruction Manual Voluson 730 English H46611E 1 N Instruction Manual Voluson 730 German H46611G 1 N Instruction Manual Voluson 730 Spanish H46611S 1 N Instruction Manual Voluson 730 Russian H46611R 1 N Instruction Manual Voluson 730 French H46611F 1 N Instruction Manual Volu...

Страница 233: ...ions Abdominal Obstetrics Gynecology with Gray Lense replaces AB2 5 KTZ195351 1 903 AB3 8 KTZ195546 electronic broadband curved array transducer frequency range of 3 8 MHz Applications Abdominal Obstetrics Gynecology Urology Pediatrics 1 904 AB4 8 KTZ195354 electronic broadband curved array transducer frequency range of 4 8 MHz Applications Abdominal Obstetrics Gynecology Pediatrics 1 905 S ACP4 7...

Страница 234: ...array transducer frequency range of 6 12 MHz electronically steerable Applications Small Parts Peripheral Vascular Pediatrics Orthopedics 1 908 SP10 16 KTZ195531 electronic broadband linear array transducer frequency range of 10 16 MHz electronically steerable Applications Small Parts Orthopedics 1 909 PA2 5 KTZ195507 electronic broadband phased array transducer frequency range of 2 5 MHz Applicat...

Страница 235: ...ventional Radiology 1 923 RAB4 8P KTZ156767 Follow up model of RAB4 8 KTZ156738 Real time 4D broadband electronic curved array transducer with a frequency range of 4 8 MHz Applications Abdominal OB Gyn Pediatrics Interventional Radiology 1 924 RSP6 12 KTZ195533 Real time 4D broadband electronic linear array transducer with a frequency range of 6 12MHz and a scan width of 40 mm Applications Small P...

Страница 236: ...obes Item Part Name Part Number Description Qty FRU 931 PCW4 0 KTZ195540 single element Continuous Wave CW Doppler pencil probe with a nominal operating frequency of 4 0 MHz no B image Applications Cardiology Pediatrics 1 932 SCW2 0 KTZ195538 single element Continuous Wave CW Doppler pencil probe with a nominal operating frequency of 2 0 MHz no B image Applications Cardiology suprasternal 1 931 93...

Страница 237: ...logy 1 943 S VDW5 8B KTZ156664 3D endocavity electronic sector transducer center frequency 6 5 MHz Applications Obstetrics Gynecology Urology 1 944 S VNA5 8B KTZ156662 3D neonatal electronic wideband transducer center frequency 6 5 MHz Applications Neonatal Cephalic Pediatrics Small parts 1 945 S VNW5 10 KTZ156695 3D broadband electronic linear array transducer with a frequency range from 5 10 MHz...

Страница 238: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL 9 34 Section 9 12 Probes This page was intentionally left blank ...

Страница 239: ...ule 10 2 10 4 Tools Required 10 5 10 5 System Periodic Maintenance 10 6 10 6 Using a Phantom 10 11 10 7 Electrical Safety Tests 10 11 10 8 When There s Too Much Leakage Current 10 24 PM INSPECTION CERTIFICATE 10 25 CAUTION Practice good ESD prevention Wear an anti static strap when handling electronic parts and even when disconnecting connecting cables DANGER THERE ARE SEVERAL PLACES ON THE BACKPL...

Страница 240: ... Testing results corrective action and the effects of corrective action must be documented and maintained on the site Your GE service representative can help you with establishing performing and maintaining records for a quality assurance program Section 10 3 Periodic Maintenance Schedule 10 3 1 How often should PMs be performed The Periodic Maintenance Schedule specifies how often your Voluson 73...

Страница 241: ...eful not to get the cloth too wet so that moisture does not enter the entire system Mechanical parts Clean and inspect the mechanical function of wheels casters brakes and swivel locks as well as side door foot rest front and rear handle and monitor holder Remove Dust and Coupling gel Mobile Unit Check Daily Control Panel movement Check Turn mechanism Mobile Unit Check Daily Trackball Check Check ...

Страница 242: ...t Checks Twice Annually Measurement Accuracy Checks Also after corrective maintenance or as required by your facilities QA program Probe Phantom Checks Check axial and lateral resolution see Basic User Manual Technical specifications Check Gain and TGC changes vary the focus and check reaction on screen Also after corrective maintenance or as required by your facilities QA program Functional Check...

Страница 243: ...ble wrist strap 3M 2214 Small adjustable wrist strap 3M 3051 conductive ground cord Anti Static Vacuum Cleaner 46 194427P278 46 194427P279 120V 230V Safety Analyzer 46 285652G1 DALE 600 KIT or equivalent for electrical tests SVHS VCR Cassette E7010GG E7010GF 60 minute 120 minute SVHS VCR Head Cleaner See VCR user manual for requirements 3 5 MOD MEDIA E8381AA E8381AB KTZ207077 blank 128 M disk blan...

Страница 244: ...Description 1 Ask Listen Ask the customer if they have any problems or questions about the equipment 2 Paperwork Fill in the top of the Periodic Maintenance PM Inspection Certificate Note all probes and system options 3 Power up Turn the system power on and verify that all fans and peripherals turn on Watch the displays during power up to verify that no warning or error messages are displayed 4 Pr...

Страница 245: ...ppler Mode operation Check the basic system controls that affect this mode of operation 5 Doppler Modes Verify basic Doppler Mode operation PW and CW if available Check the basic system controls that affect this mode of operation 6 3D Mode Verify basic 3D Mode operation Check the basic system controls that affect this mode of operation 7 Applicable Software Options Verify the basic operation of al...

Страница 246: ...he Line printer Clean heads and covers if necessary 5 DICOM Verify that DICOM is functioning properly Send an image to a DICOM device 6 Footswitch Verify that the footswitch is functioning as programed Clean as necessary 7 ECG Verify basic operation with customer Table 10 7 Mains Cable Inspection Step Item Description 1 Unplug Cord Disconnect the mains cable from the wall and system 2 Inspect Insp...

Страница 247: ...moving and release mechanism Check all wheel locks for proper operation 6 Cables Connectors Check all internal cable harnesses and connectors for wear and secure connector seating Pay special attention to footswitch assembly and probe strain or bend reliefs 7 Shielding Covers Check to ensure that all EMI shielding internal covers air flow panels and screws are in place Missing covers and hardware ...

Страница 248: ...cates the probe cannot be used Do a visual check of the probe pins and system sockets before plugging in a probe 10 5 7 3 Basic Probe Cleaning Refer to the User s Manual of the respective probe for details on cleaning NOTE To help protect yourself from blood borne diseases wear approved disposable gloves These are made of nitrile derived from vegetable starch to prevent allergic latex reactions NO...

Страница 249: ...s High leakage current can also indicate degradation of insulation and a potential for electrical failure Do not use probes or equipment having excessive leakage current To minimize the risk that a probe may shock someone the customer should Not use a probe that is cracked or damaged in any way Check probe leakage current once a year on surface probes twice a year on endocavitary probes whenever p...

Страница 250: ...ontacted mains voltage The following tests are performed at the factory and should be performed at the site These tests are grounding continuity chassis leakage current probe leakage current and ECG leakage current All measurements are made with an electrical safety analyzer Table 10 11 Chassis Leakage Current Limits Accessible Metal Surfaces Country Normal Condition Open Ground Reverse Polarity O...

Страница 251: ...ester and noting the combination of lights that are illuminated Any problems found should be reported to the hospital immediately and the receptacle should not be used NOTE No outlet tester can detect the condition where the Neutral grounded supply conductor and the Grounding protective earth conductor are reversed If later tests indicate high leakage currents this should be suspected as a possibl...

Страница 252: ...e steps to test the ground wire resistance 1 Turn the Voluson 730 unit OFF 2 Plug the unit into the meter and the meter into the tested AC wall outlet 3 Plug the black chassis cable into the meter s CHASSIS connector and attach the black chassis cable clamp to an exposed metal part of the Voluson 730 unit 4 Set the meter s FUNCTION switch to the RESISTANCE position 5 Set the meter s POLARITY switc...

Страница 253: ...e made with the unit ON and OFF with the power line polarity Normal and Reversed Record the highest reading of current When using the Microguard or a similar test instrument its power plug may be inserted into the wall outlet and the equipment under test is plugged into the receptacle on the panel of the meter This places the meter in the grounding conductor and the current flowing from the case t...

Страница 254: ...for respective test points shown in Table 10 13 8 Keep a record of the results with other hard copies of PM data kept on site Table 10 13 Chassis Leakage Current Test Condition TEST CONDITION 1 Mounting screw for probe receptacle 2 Wheel support 3 Mounting screw for CRT housing 4 Mounting screw for peripheral plugged into unit 5 Mounting screw for other peripheral powered by unit ...

Страница 255: ...ents should be made with the ground open and closed with power line polarity normal and reversed and with the ultrasound console Off and On For each combination the operating controls such as the lead switch should be operated to find the worst case condition Table 10 14 Typical Data Sheet for Chassis Source Leakage Current Unit Power Tester Polarity Switch Tester Neutral or Ground Switch Test 1 P...

Страница 256: ...wer condition combinations found in the table Record the highest leakage current measured Figure 10 4 Test Circuit for Measuring Non Isolated Patient Leads Table 10 15 Testing Power Conditions ECG Power Meter s Polarity Switch Meter s Neutral Switch ON NORM CLOSED ON NORM OPEN ON REVERSE CLOSED ON REVERSE OPEN OFF NORM CLOSED OFF NORM OPEN OFF REVERSE CLOSED OFF REVERSE OPEN POWER OUTLET H Black N...

Страница 257: ... s function selector to the LEAD ISO Select the ALL position on the lead selector Depress the rocker switch to ISO TEST to test lead isolation NOTE It is not necessary to test each lead individually or power condition combinations as required in previous tests CAUTION Line voltage is applied to the ECG leads during this test To avoid possible electric shock hazard the system being tested must not ...

Страница 258: ...imum Allowance Limit GROUND OPEN GROUND CLOSED Patient Lead to Ground Leakage Current Test and Patient Lead to Lead Leakage Current Test 115V 10uA 10uA 220 240V 500uA 10uA Table 10 17 Maximum Allowance Limit for ECG Leakage Current AC Power Source Maximum Allowance Limit Patient Lead Isolation Current Test 115V 20uA 220 240V 5mA Table 10 18 Typical Data Sheet for ECG Leakage Current ECG Power Test...

Страница 259: ...ements should be made with the ground open and closed with power line polarity normal and reversed and with the unit Off and On For each combination the probe must be active to find the worst case condition NOTE Each probe will have some amount of leakage current dependent on its design Small variations in probe leakage currents are normal from probe to probe Other variations will result from diff...

Страница 260: ...ternal probe into the meter s EXTERNAL connector 4 Set the meter s FUNCTION switch to EXTERNAL position 5 Connect the probe for test with the connector of the console 6 Add the saline probe and the imaging area of the probe into the saline bath 7 Have unit power ON for the first part turn it OFF for the second half 8 Depress the ISO TEST rocker switch and record the highest current reading 9 Follo...

Страница 261: ... Equipment damage possibility Never switch the Polarity and the status of Neutral when the unit is powered ON Be sure to turn the unit power OFF before switching them using the POLARITY switch and or the NEUTRAL switch Otherwise the unit may be damaged Table 10 19 Typical Data Sheet For Transducer Source Leakage Current Transducer Tested Unit Power Tester Power Polarity Switch Tester GROUND or NUE...

Страница 262: ... if the fault lies with the probe or the scanner NOTE Each probe will have some amount of leakage dependent on its design Small variations in probe leakage currents are normal from probe to probe Other variations will result from differences in line voltage and test lead placement The maximum allowable leakage current for body surface contact probe differs from inter cavity probe Be sure to enter ...

Страница 263: ...cy Scan Format Model Number Serial Number Probe 2 Frequency Scan Format Model Number Serial Number Probe 3 Frequency Scan Format Model Number Serial Number Probe 4 Frequency Scan Format Model Number Serial Number Probe 5 Frequency Scan Format Model Number Serial Number Probe 6 Frequency Scan Format Model Number Serial Number Probe 7 Frequency Scan Format Model Number Serial Number Probe 8 Frequenc...

Страница 264: ...orrect ground wiring config System Ground Continuity Chassis Source Leakage Current Probe Chassis Source Leakage Current Wheel Chassis Source Leakage Current CRT Patient Lead Source Leakage Lead to Ground Patient Lead Source Leakage Lead to Lead Patient Lead Source Leakage Isolation Peripheral 1 Leakage Current Peripheral 1Ground Continuity Peripheral 2 Leakage Current Peripheral 2Ground Continuit...

Страница 265: ...19 Voluson 730 5 9 Voluson 730 Basic 5 4 Board Descriptions Back End 5 26 Front End 5 21 Boot Up 3 9 4 2 C Calculation Measurements Functional Checks 4 21 General remarks to perform 4 21 CD RW Drive Theory 5 34 CE Compliance 1 10 Cine Mode Functional Checks 4 17 Color Flow Modes Functional Checks 4 12 Connectivity Worksheet 3 36 Contact Information 1 11 Control Console Rotation 5 38 Transport Lock...

Страница 266: ...Hardware Options 5 17 Hazard Icons 1 4 How to use the Auto Tester 7 5 Human Safety 1 7 I Image Management Functional Checks 4 23 Inrush Current 2 2 Installation Digital Color Printer CP770DW 3 16 Digital Color Printer UP D21MD 3 14 Digital Color Printer UP D2600S 3 16 Line Printer HP 990cxi 3 12 Line Printer HP 995c 3 12 Internal I O Theory 5 28 L Label Locations 1 8 Line Printer Installation HP 9...

Страница 267: ... of 1 1 Service Page 5 41 Shutdown Procedure 3 10 4 3 Software Options 5 15 System Components 5 3 System Configuration 3 28 How to change Language 3 29 How to enter Date and Time 3 29 How to enter Hospital Name 3 29 System Maintenance 10 6 System Manufacturer 1 12 T Theory General 5 2 Top Console 5 31 Theory 5 31 Touch Panel Functional Checks 4 4 4 5 Trackball Adjustment 6 5 Trackball top fixation...

Страница 268: ...GE MEDICAL SYSTEMS KRETZTECHNIK ULTRASOUND DIRECTION 105844 REVISION 1 VOLUSON 730 SERVICE MANUAL IV ...

Страница 269: ......

Страница 270: ... Systems Telex 3797371 P O Box 414 Milwaukee Wisconsin 53201 U S A Asia Pacific Latin America North America GE Ultraschall Tel 49 0 212 28 02 208 Deutschland GmbH Co KG Beethovenstrabe 239 Postfach 11 05 60 D 42655 Solingen German ...

Отзывы: