GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
5 - 74
Section 5-9 - Back End Processor (BEP)
5-9-15
IO Board
5-9-15-1
General description
Different I/O Boards are used for BEP6 and BEP5:
•
The BEP6 I/O Board has
two
connectors for connection to the BEP6 Motherboard.
•
The BEP5 I/O Board has
one
connector for connection to the BEP5 Motherboard.
The IO Board is interfaceing between the:
•
BEP and the connectors on the rear side of VIVID E9 (video/audio/USB)
•
BEP and the Top Console
•
BEP and the internal printer
•
Audio for the speakers, the rear plugs and the DVR
•
(BEP5 I/O ONLY) Boundary Scan from the BEP to boards in the Front End Card Rack
5-9-15-2
Location in the unit
At the rear of the BEP with the Rear panel available at the rear of the VIVID E9.
5-9-15-3
Input DC voltages
Internal in BEP from the BEP Power Supply.