Chapter 7 - Diagnostics / troubleshooting
7 - 17
7-9-3
XY Lock is not working
Follow the steps below to resolve this issue:
1.) Run the
XYZ Test Program
and select
Fix Lock
.
Section 7-8 "Motor Controller Test" on page 7-13
.
2.) If the
Lock
still does not respond when pushing the buttons in front of the UI, release the Lock
manually.
See:
4-2-3-5 "Manually releasing the XY Lock" on page 4-12
3.) Remove the
Park Lock
and adjust it manually.
6-7-3-2 "XY Lock Adjustment for Lock and Brake Mechanism" on page 6-15
7-9-4
XY Brake Motors Troubleshooting
NOTE:
Cables are not labeled at the XY controller, so right side and left side motors are not defined.
When troubleshooting the XY brake motors, keep this in mind:
•
The rear motors are #1 and #3.
•
The front motors are #2 and #4.
•
Motor #1 and #2 are on the same side.
•
Motor #3 and #4 are on the same side (opposite to motor #1 and #2).
If you are in a silent environment, it may be possible to hear which motor is activated during the test.
7-9-5
Z Movement fails
•
If it is impossible to lower the Top Console, but moving it upwards, works OK, it indicates that one
of the control switches is stuck in “Up” position.
•
If you have to help the Top Console up when moving upwards, but the motor assistance work OK
when lowering the Top Console, it indicates that the gas spring inside the Z Mechanism is failing.
-
Replace the Z Mechanism.
Содержание Vivid E9
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Страница 801: ......