Section 8-6 - Top Console Parts Replacement
8-6-9-2
Remove the Main Controller Board
Follow these steps to remove the Main Controller assembly.
1.) Lift out the Upper OP Panel/Touch Panel Assembly and place it face down on an ESD safe surface.
2.) Remove the back cover by removing the three (3x) fixing screws (Phillips #1) and the two screws.
at the D-SUB connector (3/16 inch nut driver), see:
Figure 8-125 "Operator Panel, Upper, seen from
3.) Lift (pull) the upper part of the cover so you can release it from the four protruding tabs.
NOTE:
Take care not to damage the EMC gasketing that goes around the Touch Panel Assembly,
especially during re-assembly.
Figure 8-125 Operator Panel, Upper, seen from the rear side
PULL TO REMOVE
PROTUDING TABS
(NOT VISIBLE IN THIS ILLUSTRATION)