8 - 306
Section 8-14 - Peripherals replacement
8-14-2-2
Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the VIVID E9.
2.) Disconnect the Mains Power Cable from the wall outlet.
3.) Disconnect all probes.
4.) Remove the Left Side Cover.
5.) Remove the
BEP Cover
.
a.) Loosen the BEP Cover’s fixing screw(s).
-
On BEP6 there are two finger screws on the top of the BEP Cover.
-
On newer BEP5 there are three Phillips screws, one on the top and one on each side of
the BEP Cover.
-
On older BEP5 there is one finger screw at the top of the BEP Cover.
b.) Tilt the top of the
BEP Cover
away from the
BEP
.
c.) Set the
BEP Cover
aside.
6.) For easier access to the DVR Board, remove the Hard Disk Drive.
Next (select what applies):
•
8-14-2-3 "BEP6 - Digital Video Recorder removal procedure" on page 8-307.
•
8-14-2-4 "BEP5 - Digital Video Recorder removal procedure" on page 8-308.