Chapter 8 - Replacement procedures
8 - 181
8-7-3
XY Park Lock replacement
8-7-3-1
Preparation
Unlock the XY, either by using the release knobs or by manually release the lock.
1.) At the rear of the VIVID E9, release the console’s XY mechanism by inserting a screwdriver into the
release point and pressing until release.
8-7-3-2
XY Park Lock removal
1.) Remove four screws.
2.) Carefully pull out the XY Park Lock assembly.
3.) Disconnect the cable at the connector.
4.) Remove the XY Park Lock assembly.
8-7-3-3
XY Park Lock installation
1.) Install the cable from the XY Park Lock assembly in the assigned connector.
2.) Position the XY Park Lock assembly.
3.) Install the four screws.
8-7-3-4
Verification
Verify that the XY locks and unlocks as intended.
If needed, adjust the locking mechanism as described in:
Section 6-7 "Adjusting the XYZ Mechanism" on page 6-12
.
Figure 8-187 XY mechanism release
Figure 8-188 Remove four screws
XY
(FROGLEG)
RELEASE