GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
8 - 180
Section 8-7 - Replacing XYZ Parts
8-7-2-3
XY-Brake Assy installation procedure
NOTE:
Be careful when fighting screws that are entered into plastic. Use 2-3 Nm. If you tighten to
much, the internal screw-threads in the plastic are destroyed, and the plastic part must be
replaced.
1.) Position the XY-Brake Assy into the arm. You may need to turn the axle on the motor (by hand) to
adjust the lever so it fits.
2.) Install the fixing screw for the XY Brake Assy.
3.) Connect the cable from the motor to the XY Brake cable.
4.) Position the Cover so it aligns with the hole(s) for the fixing screw(s).
5.) Install the fixing screws:
-
The covers for the front legs are fixed with one screw. Use low Torque, max 2.5 Nm (Plastic
part)
-
The covers for the rear legs are fixed with two screws. Use low Torque, max 2.5 Nm (Plastic
part)
Содержание Vivid E9
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Страница 801: ......