Section 4-3 - Functional checks
4-3-4
Color Mode Checks
4-3-4-1
Introduction
Color Flow screens are 2D or M Mode screens with colors representing blood or tissue movement.
Color Flow may be selected both from 2D mode or from M mode or a combination of these.
4-3-4-2
Preparations
1.) Connect one of the probes to the scanner’s left-most probe connector.
-
See
"Connecting Probes" on page 3-24
for info about connecting the probes.
2.) Turn ON the scanner.
The 2D Mode window is displayed (default mode).
4-3-4-3
Select Color 2D Mode
1.) From an optimized 2D image, press
Color.
2.) Use the trackball (assigned function: Pos) to position the ROI frame over the area to be examined.
3.) Press
Select
. The instruction
Size
should be highlighted in the trackball status bar.
NOTE:
If the trackball control Pointer is selected, press trackball to be able to select between Position
and Size controls.
4.) Use the trackball to adjust the dimension of the ROI.
4-3-4-4
Adjust the Color 2D Mode controls
•
Adjust the
Active mode gain
to set the gain in the color flow area.
•
Adjust
Scale
to the highest setting that provides adequate flow detection.
NOTE:
The scale value may affect FPS, Low Velocity Reject, and Sample Volume.
•
Adjust
Low Velocity Reject
to remove low velocity blood flow and tissue movement that reduces
image quality.
•
Adjust
Variance
to detect flow disturbances.
•
Adjust
Sample volume
(SV) to a low setting for better flow resolution, or a higher setting to more
easily locate disturbed flows
•
Adjust
Frequency
to optimize the color flow display. Higher settings improve resolution. Lower
settings improve depth penetration and sensitivity. This does not affect the frequency used for 2D
and M-Mode.
NOTE:
NOTE: Frequency setting may affect FPS, SV and Low Velocity Reject.
•
Adjust
Power
to obtain an acceptable image using the lowest setting possible.
NOTE:
The Power setting affects all other operating modes.
Содержание Vivid E9
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Страница 801: ......