GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
8 - 282
Section 8-11 - Back End Processor (BEP) parts replacement
8-11-9
Patient I/O Interface Cable replacement
The Patient I/O Interface Cable connects the Patient IO to the BEP’s motherboard.
It is included in the BEP6.X Cable Kit - Spare Part, Part Number 5433408-80.
8-11-9-1
Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the VIVID E9.
2.) Disconnect the Mains Power Cable from the wall outlet.
3.) Disconnect all probes.
4.) Remove the Left Side Cover.
5.) Remove the
BEP Cover
.
a.) Loosen the BEP Cover’s fixing screw(s).
-
On BEP6 there are two finger screws on the top of the BEP Cover.
-
On newer BEP5 there are three Phillips screws, one on the top and one on each side of
the BEP Cover.
-
On older BEP5 there is one finger screw at the top of the BEP Cover.
b.) Tilt the top of the
BEP Cover
away from the
BEP
.
c.) Set the
BEP Cover
aside.
6.)
8-11-9-2
Patient I/O Interface Cable removal
Follow these steps to remove the
Patient I/O Interface Cable
:
1.) Remove the
BEP Power Board
. This is required to get access to the
Patient I/O Interface Cable
.
2.) Unplug the two connectors (Power and USB) from the Patient I/O.
3.) Unplug the connector from
CN USB11
on the BEP’s motherboard.
Содержание Vivid E9
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Страница 801: ......