GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 8 - Replacement procedures
8 - 117
8-6-11-2
High Voltage Inverter Board with Cable removal procedure
1.) Disconnect the cables to the High Voltage Inverter Board.
2.) Remove the two fixing screws.
3.) Remove the High Voltage Inverter Board with Cable.
8-6-11-3
High Voltage Inverter Board with Cable installation procedure
1.) Align the board’s fixing holes to the respective holes on the Main Controller Board.
2.) Install the two fixing screws. (Size: M 2.5)
3.) Plug in the cable to the connector on the Main Controller Board.
4.) Install the Operator Panel, Upper.
5.) Install the Control Knobs.
Figure 8-131 Remove cables
Figure 8-132 Remove screws
Содержание Vivid E9
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Страница 801: ......