Chapter 4 - General procedures and Functional checks
4 - 39
4-3-3
M Mode Checks
4-3-3-1
Introduction
4-3-3-2
Preparations
1.) Connect one of the probes, to the scanner’s left-most probe connector.
-
See
3-7-6 "Connecting Probes" on page 3-24
for info about connecting the probes.
2.) Turn ON the scanner.
The 2D Mode window is displayed (default mode).
3.) Press
MM
on the Operator panel to bring up an M-Mode picture on the screen.
4.) Use the trackball to position the cursor over the required area of the image.
4-3-3-3
Adjust the M Mode controls
These Image Controls are used to optimize the M mode picture. Verify that all the listed controls are
working as intended:
•
Adjust
Horizontal sweep
to optimize the display resolution.
•
Adjust
Gain
and
TGC
controls to adjust the range to be imaged.
•
Use the
Frequency
(move to higher frequencies) or the
Frame rate
control (move to lower frame
rate) to increase resolution in image.
•
Use the
Frequency
(move to lower frequency) to increase penetration.
•
Adjust
Focus
to move the focal point(s) around the region of interest in the M-Mode display.
•
Adjust
Dynamic range
to optimize the useful range of incoming echoes to the available grey scale.
•
Adjust
Compress
and
Edge Enhance
to further optimize the display.
•
Adjust
Reject
to reduce noise while taking care not to eliminate significant low-level diagnostic
information.
Figure 4-38 M-Mode Screen Example
Содержание Vivid E9
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Страница 801: ......