GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 8 - Replacement procedures
8 - 141
8-6-22
Operator Panel Cable Kit Replacement
8-6-22-1
Overview
The Operator Panel Cable Kit includes three different cables, as illustrated in
•
The Trackball USB Cable is used on the Operator Panel, Lower.
•
The USB Video Board Flex Cable and the HV Inverter Cable are both used on the Operator Panel,
Upper.
Figure 8-153 The Operator Panel Cable Kit
TRACKBALL USB CABLE
Connects the Trackball to the Controller board.
USB VIDEO BOARD FLEX CABLE
Connects the Touch Screen to the USB Video board.
HV INVERTER CABLE
Connects the Touch Screen to the HV Inverter board.