Follow these steps to install the BEP6 Power Board Assembly:
1.) Position the BEP6 Power Board Assembly so it enters the two guides and slide the BEP6 Power
Board Assembly in so it is plugged into the BEP6’s backplane. You may need to apply a pressure
to make it slide into position.
2.) Install the two screws that secure the BEP6 Power Board Assembly to the BEP frame.
3.) Plug in the cable plug in the connector near the top of the BEP6 Power Board Assembly.
4.) Install the BEP Cover as described in these steps:
a.) Insert the base of the BEP Cover inside the base of the BEP.
Be sure the bottom lip of the BEP Cover rests inside the BEP.
b.) Tilt the top of the BEP Cover toward the VIVID E9.
c.) Install the fixing screw(s) for the BEP Cover:
-
On BEP6 there are two finger screws on the top of the BEP Cover.
-
On newer BEP5 there are three Phillips screws, one on the top and one on each side of
the BEP Cover.
-
On older BEP5 there is one finger screw at the top of the BEP Cover.
5.) Replace the Left Side Cover.
NOTICE
Be careful not to pinch any of the cables when installing the BEP Cover.
Содержание Vivid E9
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Страница 801: ......