GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
5 - 90
Section 5-11 - Input and Output (I/O) modules
5-11-2-3
Input DC Voltages
+5 VDC
+12 VDC
Pinout for the DC input on the Patient I/O module
5-11-2-4
Patient I/O - inputs
•
ECG / Respiration
•
Phono (from a phono heart microphone)
•
AUX - Analog Input (Pulse/Pressure)
Pinout for the AUX connector
The pinout for the AUX connector is described in the table below:
The AUX is default a 1 Vpp (Volt peak-to-peak) input with a max frequency of 300 Hz. It has a
programmable high gain mode with a maximum input signal of 300 mVpp (millivolt peak-to-peak).
The inputs are differential. For a single ended sensor signal, the pin 1 (input -) should be connected to
the GND of the sensor.
Table 5-25 DC input pinout on the Patient I/O Module
CONNECTOR
SIGNAL NAME
DC INPUT
•
Pin 1: +12VDC
•
Pin 2: GND
•
Pin 3: GND
•
Pin 4: +5VDC
Table 5-26 Pinout for the AUX Connectors
CONNECTOR
SIGNAL NAME
AUX
•
Pin 1: Input -
•
Pin 2: Input +
•
Pin 3: Gnd
•
Pin 4: Nasal Sensor 1
•
Pin 5: Nasal Sensor 2
4
3 2
1
1
2
3
4
5