Section 8-14 - Peripherals replacement
8-14-6-3 Install the GFI PCIe Cable (cont’d)
4.) Connect the other end of the GFI PCIe Cable to the connector on top of the BEP.
Ensure that the GFI PCIe Cable is kept out of the drawed box area in
, as there is a
risk that the Z Mechanism damage the cable.
5.) Install the
Left Side Cover and the Right side Cover
.
Figure 8-259 Keep cables out of the marked area
1. GFI PCIe Cable (at top of BEP)
2.
Keep cables out of the marked area