Install the Lower Bezel
1.) Install the Switch Cups in the correct positions on the Lower Bezel. Use the mirrored image in
as a reference.
NOTE:
THE ILLUSTRATION BELOW IS A MIRRORED ILLUSTRATION.
USE IT FOR REFERENCE WHEN INSTALLING THE SWITCH CUPS!
2.) Install the Lower Switch Board with Elastomer.
3.) Install the Encoder Board.
4.) Install the Trackball.
5.) Install the Operator Panel, Lower.
6.) Install the Operator Panel, Upper.
7.) Install the Knobs for Encoders and Slidepots.
Figure 8-149 Mirrored illustration of the Switch Cups positions