![GE Vivid E9 Скачать руководство пользователя страница 238](http://html1.mh-extra.com/html/ge/vivid-e9/vivid-e9_service-manual_4205051238.webp)
GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
5 - 48
Section 5-8 - Front End Processor (FEP)
5-8-8-2
Location in the Unit
The GFI2 board is plugged into the Back Plane as the right most board in the FE rack.
5-8-8-3
Input DC Voltages
•
+24 DC (from Main LV Power via FEP Backplane)
•
+/- 15 VDC
•
+/- 6VA DC
•
BEP5V
Other needed voltages are generated locally on the GFI.
5-8-8-4
Clocks
These clocks are generated on the board:
•
200 MHz for on board (GFI) use
•
200/50 MHz (for the RX board)
•
200 MHz for the RX board
Figure 5-34 GFI2 location
GFI2 BOARD