Section 8-6 - Top Console Parts Replacement
2.) Unscrew the fixing screw on the rear side of the LCD Monitor assembly.
3.) Remove the Monitor Cables Cover. Carefully insert a screwdriver into the Removal Slot” to separate
the cover from the LCD Monitor assembly.
4.) Lift the cover away and place it on a safe place.
Figure 8-85 LCD Monitor assembly- rear view
MONITOR
CABLES
COVER
FIXING SCREW
LCD MONITOR ASSEMBLY
TILTED FORWARDS
REMOVAL SLOT