GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 8 - Replacement procedures
8 - 265
8-11-4-4
Removing the Graphics Adapter from the BEP6
1.) Remove the HDD to get easier access in the rest of the procedure.
2.) Unscrew the fixing screws from the connector at top of the Graphics Adapter.
-
If DVR is installed: the flex between the Graphics Adapter and the DVR board uses finger
screws.
-
Without DVR: the flex from the motherboard is fixed with Hex screws.
3.) Disconnect the plug.
4.) Remove the fixing screw for the Graphics Adapter.
5.) Pull out the Graphics Adapter.
Figure 8-235 Graphics Adapter position in BEP6
GRAPHICS ADAPTER