8-6-5-2 Remove the LCD Cables (cont’d)
2.) Remove the bumper cover from the end of the first arm, as shown in Figure 8-99.
3.) Remove the covers from all three arm sections of the arm assembly.
a.) Remove the M4 x 8 screw holding the first arm section cover.
b.) Remove the M4 x 8 screw holding the third arm section cover.
Figure 8-99 Bumper Cover, Removed