8 - 90
Section 8-6 - Top Console Parts Replacement
8-6-5-2
Remove the LCD Cables
Follow these steps to remove the LCD Cables:
NOTE:
When handling the arm, use the arm lock unless the arm is being turned. Locking will help
stabilize the arm.
NOTE:
LCD cable color may be different than pictured.
1.) Remove the two M4 x 8 screws holding the Bumper Cover to the Arm.
Figure 8-97 Three Arm Sections
Figure 8-98 Screws for Bumper cover
First Arm
Section
Second Arm
Section
Third Arm
Section
Bumper Cover