GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 4 - General procedures and Functional checks
4 - 23
4-2-7-6
Connectivity — Recording the TCP/IP settings
1.) Select
Config/F2
to log on as
ADM
. (Default password: ulsadm).
For a detailed description, see:
4-2-5 "Logging on to VIVID E9 as ‘ADM’" on page 4-15
2.) Select
Connectivity
>
Tcpip
.
3.) Record all settings in:
Table 4-1 "Record settings from Tcpip screen and sub-screens" on page 4-24
.
4.) Select
Network Settings
.
5.) Double-click
Local Area Connection
.
6.) Select
Properties
.
7.) Double-click
Internet Protocol (TCP/IP)
.
8.) Select
Advanced
9.) Record the settings in
Table 4-1 "Record settings from Tcpip screen and sub-screens" on page 4-24
.
-
Record if
DHCP
is Enabled.
-
Record
IP address
,
Subnet mask
and
Default gateways
Figure 4-22 Tcpip screen, Example Settings
Figure 4-23 Tcpip > Network Settings
NETWORK SETTINGS