GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
Chapter 8 - Replacement procedures
8 - 281
8-11-8
SATA Jumper replacement
The
SATA Jumper
connects
CN DV11
to
CN DV12
in 2D systems wilhout DVR.
The jumper is included in the
BEP6.X Cable Kit - Spare Part
, Part number 5433408-80.
8-11-8-1
Preparations
When preparing for the replacement, you must perform the following steps:
1.) Power down the VIVID E9.
2.) Disconnect the
Mains Power Cable
from the wall outlet.
3.) Disconnect all probes.
4.) Remove the
Left Side Cover
.
5.) Remove the
BEP Cover
.
a.) Loosen the BEP Cover’s fixing screw(s).
-
On BEP6 there are two finger screws on the top of the BEP Cover.
-
On newer BEP5 there are three Phillips screws, one on the top and one on each side of
the BEP Cover.
-
On older BEP5 there is one finger screw at the top of the BEP Cover.
b.) Tilt the top of the
BEP Cover
away from the
BEP
.
c.) Set the
BEP Cover
aside.
6.)
8-11-8-2
SATA Jumper removal
Disconnect the
SATA Jumper
from the connectors
CN DV11
and
CN DV12
on the motherboard.
8-11-8-3
SATA Jumper installation
1.) Install the
SATA Jumper
in the connectors
CN DV11
and
CN DV12
on the motherboard.
2.) Install the
BEP Cover
.
3.) Install the
Left Side Cover
.
Содержание Vivid E9
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Страница 801: ......