GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
8 - 22
Section 8-4 - Reloading the software
8-4-10-5 Press A. Install System Software. (Erase all partititions) (cont’d)
The BEP starts to copy the Image file from the DVD to the Repository partition on the HDD. This will
take typically eight to ten minutes.
Next, the copied Image file is used as source when the System software is installed on the C:\ partition.
Several screens will be displayed.
When the installation of the System Software has finished, the DVD is ejected and you are requested
to remove the System Software DVD and insert the Application Software CD.
4.) Continue with:
"Loading the Application Software" on page 8-23
Figure 8-11 Move contents from DVD to Repository
Figure 8-12 Installing the System Software
Saving Imagefile to repository (takes up to 10 minutes).
Please wait.
Содержание Vivid E9
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Страница 801: ......