![GE Vivid E9 Скачать руководство пользователя страница 772](http://html1.mh-extra.com/html/ge/vivid-e9/vivid-e9_service-manual_4205051772.webp)
GE H
EALTHCARE
D
IRECTION
GA091568, R
EVISION
5
VIVID E9 S
ERVICE
M
ANUAL
10 - 16
Section 10-5 - System maintenance
10-5-6-5
Cleaning TEE probes
To clean the handle:
•
Use a cloth or towel only lightly moistened with cleaner/ disinfectant (as recommended in TEE
Probe Care Card).
Avoid too much moisture on the cloth/ towel.
•
To remove residual chemicals wipe the handle with a cloth or towel lightly moistened with clean
water.
•
Dry with clean cloth or towel.
To clean the cable:
•
Use a cloth or towel only lightly moistened with cleaner/ disinfectant (as recommended in TEE
Probe Care Card).
•
Avoid too much moisture on the cloth/ towel as this may cause cleaning fluids to flow along the cable
and into the connector/ handle.
•
Always clean the cable from the connector towards the middle of the cable.
•
Hold the connector higher than the cable to avoid that cleaning fluids flows along the cable and into
the connector.
•
Use the same technique when cleaning the cable from the handle side.
•
To remove residual chemicals follow the same prosedure as above but use a cloth or towel lightly
moistened with clean water.
•
Wipe off moisture with a cloth or towel and place the probe in a TEE Storage Rack and let it dry for
30 min.
Figure 10-1 Cleaning cable
Содержание Vivid E9
Страница 1: ...GE Healthcare OPERATING DOCUMENTATION VIVID E9 Service Manual Part Number GA091568 Revision 5...
Страница 2: ......
Страница 11: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL ix ZH CN KO...
Страница 34: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL xxxii...
Страница 800: ...GE HEALTHCARE DIRECTION GA091568 REVISION 5 VIVID E9 SERVICE MANUAL Index 12...
Страница 801: ......