Section 8-14 - Peripherals replacement
8-14-2-4
BEP5 - Digital Video Recorder removal procedure
Follow these steps to remove the Digital Video Recorder from a BEP5:
1.) Locate the DVR Board behind the HDD.
2.) Disconnect the audio jack, SATA, DVI flex cables where they attach to the DVR Board.
3.) Remove the DVR Board from the motherboard slot.
Next:
•
8-14-2-6 "BEP5 - Digital Video Recorder installation procedure" on page 8-311.
Figure 8-255 DVR Board (BEP5)
Figure 8-256 DVR Cables (BEP5)
DVR Board
BEP HDD HINGES
(HDD REMOVED)
SATA
AUDIO
JACK
DVI FLEX