GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Section 8-7 - Mechanical Option Installation instruction
8 - 155
8-7-1
Flexible LCD Arm installation Procedure
8-7-1-1
Tools
•
Common pilIips screwdrivers
•
Allen/Unbraco wrench
•
Stubby screwdriver (Flat tip and Cross tip)
8-7-1-2
Preparations
•
Shut down the system and switch off the main breaker.
•
Maneuver control console to a suitable position for removing the monitor.
8-7-1-3
Removal procedure
1.) Remove the LCD cable cover. Refer to the figure below.
2.) Unscrew 1 screw (2306565, BH M4x16 WHT) to remove the power cord bracket assembled under
the LCD cable cover. Refer to the figure below.
Figure 8-231 Removing the LCD cable cover
Figure 8-232 Removing the power cord bracket
Содержание LOGIQ P6
Страница 2: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 2 ...
Страница 9: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SEVICE MANUAL vii JA ZH CN KO ...
Страница 22: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 10 ...
Страница 130: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL 4 50 Section 4 9 Board Function Checks ...
Страница 477: ......