GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
8 - 96
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-45
AP950 Assy
8-2-45-1
Tools
•
Common philips screwdrivers
•
Allen/Unbraco wrench
8-2-45-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-45-3
Removal procedure
-
To detache the AC Power assy, All open 4 pcs console cover that are left, right, front, rear
cover.
1.) Remove side left cover and right side cover. Refer to the
8-2-14 "Right or Left Side Cover" on page
8-2-14 "Right or Left Side Cover" on page 8-47
.
2.) Remove rear side cover and front cover. Refet to the
8-2-15 "Front Cover" on page 8-49
and
.
3.) Unscrew 2 M4 screws and 2 Hexa screws on the front of AC trans box.Refer to the Figure 8-141.
4.) Loosen 2 Hex (left side in Figure 8-144)screws and 4 M4 screws (right side of Figure 8-142).
Figure 8-141 4 screw on the front of the AC TRANSBOX
Figure 8-142 Back of the AC TRANSBOX
Содержание LOGIQ P6
Страница 2: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 2 ...
Страница 9: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SEVICE MANUAL vii JA ZH CN KO ...
Страница 22: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 10 ...
Страница 130: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL 4 50 Section 4 9 Board Function Checks ...
Страница 477: ......