4 - 28
Section 4-7 - Keyboard Function Checks
4-7-6
TGC key assy / TGC Knob Set validation
1.) Follow up the procedure
4-7-1 "Main Keyboard Assy Function Validation Procedure" on page 4-24
to run the scfpapitest program.
2.) Slide TGC pot and check up the value.
4-7-7
Sub Keyboard Encoder Knob Set validation
4-7-2 "Sub Keyboard Assy Function Validation Procedure" on page 4-27
NOTICE
While sliding TGC pot, TGC message should be displayed.