GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
8 - 88
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-41
NEST FAN Assy
8-2-41-1
Tools
•
Common philips screwdrivers
8-2-41-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-41-3
Removal procedure
1.) Remove the Left and Right Covers. Refer to the
8-2-14 "Right or Left Side Cover" on page 8-47
2.) Remove the EMI Right and Left covers. Refer to the
8-2-24 "EMI Cover L and R" on page 8-62
3.) Cut the cable tie, and unplug the ‘FAN’ connector. Refer to the Figure 8-131.
4.) Unscrews the two screws. Refer to the Figure 8-132.
5.) Pull out the FAN assy forward from the nest. Refer to the Figure 8-132.
Figure 8-131 Connector and Screw2 screws on the Nest Fan
Figure 8-132 Pull out the nest fan