GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
8 - 12
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-4
New Articulation arm with cover installation Procedure
8-2-4-1
Tools
•
Common philips screwdrivers
•
Allen/Unbraco wrench
•
Stubby screwdriver (Flat tip and Cross tip)
8-2-4-2
Preparations
•
Shut down the system and switch off the main breaker.
•
Maneuver control console to a suitable position for removing the monitor.
8-2-4-3
Removal procedure
1.) Remove the LCD cable cover. Refer to the figure below.
2.) Unscrew 1 screw (2306565, BH M4x16 WHT) to remove the power cord bracket assembled under
the LCD cable cover. Refer to the figure below.
Figure 8-17 Removing the LCD cable cover
Figure 8-18 Removing the power cord bracket