Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8 - 91
5.) Disconnect three (3) USB connectors or 5 USB connectors from the rear panel and 2 AC power
connector. The 4D controller USB and ECG USB connector also connect to this rear paenl PWA.
Refer to the Figure 8-135.
6.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-2-42-4
Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-135 Open Rear Panel
Table 8-43 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test
Debrief Script
“Service Manual, Direction
5245279, Rev 3+, Section 8-2-42. Equipment
passed all required tests and is ready for use. “
ARPII Assy Function Validation Procedure