GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
8 - 30
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
24.)Screw 2 screws (2337572, FH M3X6 WHT) to assemble the Arm cover L & R. Refer to the figure
below.
25.)Screw 4 screws (2159634, BH M4X10 WHT) to assemble the Arm bottom Cover. Refer to the figure
below.
Figure 8-59 Assembling the Arm Cover L & R
Figure 8-60 Assembling the Arm bottom cover
Содержание LOGIQ P6
Страница 2: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 2 ...
Страница 9: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SEVICE MANUAL vii JA ZH CN KO ...
Страница 22: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 10 ...
Страница 130: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL 4 50 Section 4 9 Board Function Checks ...
Страница 477: ......