Section 7-9 - Trouble Shooting Tree
7 - 27
Section 7-9Trouble Shooting Tree
7-9-1
System Does Not Power On / Boot Up
Figure 7-13 Power On / Boot Up Flowchart
Yes
Is the scanner
plugged in?
Is the AC at the
wall outlet?
Make sure the
scanner is not
plugged into
emergency power
(red outlet);
Is the circuit
breaker ON?
Yes
Press the on/off switch to
boot the system. Is the
on/off switch illuminated
with an green color and
can you listen fan sound
of inside?
Yes
Yes
Does the scanner
boot up?
Yes
Scanner is ON and ready for use.
NO
NO
NO
NO
NO
Check if wall power is alive
Check if the standby 5V od APS is alive
Make sure AC power connector and con-
trol signal connector is plugged properly
If HDD indicator is blinking after power
on, check the monitor cable
Check BIOS or replace PM SOM module