GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
8 - 32
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-5
New Articulation cover set installation Procedure
8-2-5-1
Tools
•
Common philips screwdrivers
•
Stubby screwdriver (Flat tip and Cross tip)
8-2-5-2
Preparations
•
Shut down the system and switch off the main breaker.
•
Maneuver control console to a suitable position for removing the monitor.
8-2-5-3
Removal procedure
1.) ARM COVER L & R
Unscrew 2 screws (2337572, FH M3x6 WHT) to remove the Arm cover L & R. Refer to the figure below.
2.) ARM CAM COVER L & R
Unscrew 2 screws (2159632, BH M4x6 WHT) to remove the Arm cam cover L & R. Refer to the figure
below.
Figure 8-62 Removing the Arm cover L & Arm cover R
Figure 8-63 Removing the Arm cam cover L & Arm cam cover R