7 - 30
Section 7-9 - Trouble Shooting Tree
Figure 7-16 System does not Boot (contd.)
System doesn’t boot! (cont’d)
1.) Check the Control
Panel cable
connection.
2.) Check/replace
Control Panel.
YES
YES
NO
Are the Control Panel
keys lit?
Is regular scan display
shown on the monitor?
Press and hold the main switch for
five (5) seconds to reboot the
system.
NO
Go to Scan Procedure section.
END