Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-43
SATA HDD Assy
8-2-43-1
Tools
•
Common philips screwdrivers
•
Cutter
8-2-43-2
Preparations
•
Shut down the system and switch off the main breaker.
8-2-43-3
Removal procedure
1.) Remove the side left cover. Refer to the
8-2-14 "Right or Left Side Cover" on page 8-47
.
2.) Remove EMI L Cover. Refer to the
8-2-24 "EMI Cover L and R" on page 8-62
3.) Disconnect HDD power cable and unscrew 4 screws of the HDD fixture bracket. Refer to the Figure
4.) Disconnect the SATA cable from the backplane on the bottom of backplane.
NOTE:
Push inside of latch that is ejector button on the SATA connector, and pull out the cable
smoothly. This connector is a locking type.
Figure 8-136 HDD in System
Figure 8-137 Disconnect the SATA Cable