GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8 - 73
6.) Mount ASIR ROM to new SYSCONML Assy. Refer to the Figure 8-115.
7.) Confirm mounting direction and pin number.
NOTE:
It is important to swap ASIR ROM(U23). Because it contains the serial Number and option the
system.
Figure 8-114 Seperation of ASIR ROM
Figure 8-115 ASIR ROM mounting direction