GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
Section 4-7 - Keyboard Function Checks
4 - 25
4-7-1 Main Keyboard Assy Function Validation Procedure
6.) Write the 'scfpapitest.exe' on the command prompt and press 'Enter' button to pop up the
scfpapitest program.
7.) Press the keyboard button and rotate rotary button. When you press or rotary keys, Log dialog box
displays the key log. Button name and displaying log name is like the table.
Figure 4-28 scfpapitest program
Table 4-7
Button and Rotaty Table
Button or Rotary Name The Name displayed in Log
Button or Rotary Name
The Name displayed in Log
Patient
New Patient Button
Active Mode
Active Mode Button
Probe/Preset
Probe Button
CF
CFM PushEncoder Button
Report
Report Button
PDI
PDI Button
End Exam
End Exam Button
Dept (Button)
Depth PushEncoder Button
Record
Record button
Dept (Rotary)
Depth PushEncoder Rotary
P1
Print or record button1
PW
PWD PushEncoder Button
P2
Print or record button2
CW
CW Button
Reverse
Reverse Button
M
M Mode PushEncoder
Userdefine 1
Userdefine 1 Button
Scan Area
ScanArea Button
Userdefine 2
Userdefine 2 Button
M/D Cursor
MD Cursor Button
Userdefine 3
Userdefine 3 Button
Zoom (Button)
Zoom/Size PushEncoder Button
Userdefine 4
Userdefine 4 Button
Zoom (Rotary)
Zoom/Size PushEncoder Rotary
Userdefine 5
Userdefine 5 Button
Measure
Measure Button
L (Left)
Left Button
BodyPattern/Ellipse (Button)
BodyPattern/Ellipse PushEncoder
Button
Содержание LOGIQ P6
Страница 2: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 2 ...
Страница 9: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SEVICE MANUAL vii JA ZH CN KO ...
Страница 22: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 10 ...
Страница 130: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL 4 50 Section 4 9 Board Function Checks ...
Страница 477: ......