GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
7 - 20
Section 7-7 - Common Diagnostics
7-7-4
FRU Test
7-7-4-1
SYSCONML
These programs are provided for testing the SYSCONPM/SYSCONCM board.
Table 7-2
FRU Test, SYSCONPM/SYSCONCMDiagnostics Menu
Menu
Descriptions
User
Level
Tools
Full Test
Performs all of the SYSCONML assy tests listed below
All
None
Assy Revision Test
Test EEPROM.
Send test pattern to write to EEPROM and read
Displays the board revision of the SYSCONML assy
All
None
USC I/F Test
The DTRX64II, PI128II, optional CWD64 assya are connected to the SYSCONML via
the FEBUS.
The DSP in SYSCONML accesses the boards above to check if the FEBUS is
connected properly to each assy.
All
None
FEBUS Test
Checks if The DSP in SYSCONML accesses the assys, DTRX64II and CWD64, via
the FEBUS.
All
None
TRIG Test
Check the trigger max, min and ever interval
All
None
PCI I/F Test
Sends the command to the DSP to check the SDRAM in SYSCONML
All
None
DSP Memory Test
Sends the command to the DSP to check the internal DSP memory
All
None
Board Temperature
Read SYSCONML Assy Temperature.
All
None
CHACOM I/F Test
Sends the command to read / write data from DSP to CHACOMs
All
None
Содержание LOGIQ P6
Страница 2: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 2 ...
Страница 9: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SEVICE MANUAL vii JA ZH CN KO ...
Страница 22: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL Page 1 10 ...
Страница 130: ...GE HEALTHCARE DIRECTION 5245279 REVISION 3 LOGIQ P6 P6 PRO SERVICE MANUAL 4 50 Section 4 9 Board Function Checks ...
Страница 477: ......