Section 8-4 - Software Installation
8 - 121
8-4-2-4
DMC Checks (DMC User Only)
NOTE:
DMC Cable is supplied only as an option.
1.) Check DMC Cable (USB to Serial Bridge).
2.) Plug in the Cable as shown on the right side of the USB port. Refer to the Figure 8-180.
3.) Ensure
USB to Serial Bridge port
is recognized by the system.
4.) Trackball to My computer icon and left click
5.) Right click on
Properties
.
6.) Right click on the
Hardware
tab.
7.) Right click on the device manager in the middle right of the hardware tab window.
8.) Make sure
“USB to Serial Bridge”
is displayed as shown
Figure 8-180 DMC Cable
Figure 8-181 Device Manager Windows
DMC Cable
Plug on the right