GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
7 - 38
Section 7-9 - Trouble Shooting Tree
Figure 7-23 Printer Troubleshooting (contd.)
Still unable to use the Printer(s)!
Check the USB cables Between
backplane and usb port.
YES
NO
Is printing unavailable
even by manual
operation?
Is there an error message
on the monitor display?
Replace SOM module.
NO
YES
Replace the Printer.