GE H
EALTHCARE
D
IRECTION
5245279, R
EVISION
3
LOGIQ™ P6/P6 P
RO
S
ERVICE
M
ANUAL
8 - 100
Section 8-2 - DISASSEMBLY/RE-ASSEMBLY
8-2-47
Read and Write Serial Info
8-2-47-1
How to start serial no read/write program
1.) Plug the service dongle on the usb port on rear panel.
2.) Turn on the system.
3.) When the Start Application dialog box appears, press [F2] before it automatically starts
Echoloader.)
NOTE:
If maintenance access box does not appear and the scanning mode starts, just insert the
Service Dongle and press “CTRL+ALT+R”.
4.) When the Maintenance Access dialog box pops up, type in the service password.
Figure 8-147 Start Appllication Dialog Box
Figure 8-148 Maintenance Access Dialog Box